background image

SETTINGS PAGE Access Pop-up

Pin:

 5 digit number

If created during initial setup enter Pin number, 

otherwize, leave blank and click Open.

SETTINGS PAGE

--Wi-Fi Setting--

These fields are populated from the initial 

SmartLOG SETUP. 

To change network...

Network Name:

 SSID (Service Set Identifier) 

is the name of your wireless network

Password:

 network password.

Username: 

is for Enterprise Networks.

--Logger Configuration--

Access Pin:

 5 digit number (required to save)

Logger Name:

 The name is used to access the 

Dashboard and also for Email & SMS messages 

sent. Do not use the same name if using 

additional data loggers on the same network. 12 

alphanumeric characters, hyphen “-”, no spaces.

Facility Name:, Location:, Pin or ID#:

are optional fields that show on reports.

P1 & P2 Name:

 A drop down list to select Fridge 

or Freeze that will auto populate 

Alarm Temp:

 

min and max values. Custom name, min/max 

values can be entered. For single sensor set up, 

leave P2 fields empty.

Sensor SN#:

 Enter serial number of orginal 

wireless sensor or new replacement sensor. 

Cert Expires:

 Add 2 years from in service date. 

Scale: 

Display & Report in °C or °F.

Log Interval:

 

Select from 1 to 60 minutes.

Time Zone: 

Set your local Time Zone. 

Time is automatically synchronized from internert.

Note: All times are in Standard Time, no DST.

--Email From Acct Info-- 

 

see note ‘Office 365 accounts’

Email:

 Sign in to Email for 

sending

 alerts

Password:

 Email account password

Outgoing SMTP:

 Dropdown list of popular email 

provider servers. Selecting a server name will 

auto populate the 

Port #:

Custom server address & port # can be entered.

If Alerts and Status and message are not needed, 

leave Email From Acct fields empty.

--Email / SMS Contacts--

Email:

 Address to send alert email to.

SMS:

 10 digit phone number to send text msg to.

Carrier:

 Select carrier of cell phone #.

Select from dropdown list of popular cell phone 

carriers or enter custom value.

--Daily Status Messages--

Time:

 can be set to OFF for both am and pm to 

disable status messages.

Select one or both am or pm for desired 

notification times.

SPECIFICATIONS:

SmartLOG® DDL Monitor Gateway

Wi-Fi: 

802.11 b/g/n (2.4 GHz only) 

Security:

WPA/WPA2 Enterprise

Display:

240x320 pixel TFT Color Display w/ Touch Screen

Screen Update:

60 seconds 

Memory:

16MB (2 years of records @ 5 minute intervals)

Ambient Sensor:

Temperature range -40 to +125°C (-40°F to 257°F)

Humidity range 0-100%RH

Alarms:

Audible Piezo Buzzer / Visual LED / SMS / Email

Date/Time:

NTP clock synchronization during boot

Power:

5v wall adapter  w/ 5’ micro USB cable

Batteries:

Alkaline 1.5v 3 x AA (backup backup only for power loss)

Battery Life:

~ 3 days with no alarms 

(NOTE: replace all batteries after any power loss of more than 1 day)

Working Temp:

-20 °C ~ 70 °C

Dimensions:

83 (W) X 120 (H) X 26 (D) mm

Wireless Buffered Temperature Sensor

Temperature range:

-40C to 50°C (-40°F to 122°F)

Accuracy:

±0.5°C (±1.0°F)

Distance:

50m/164ft (open field), 10m/32ft (inside fridge/freezer)

Buffer medium:

Glycol

Batteries:

Alkaline 1.5v 2 x AAA

Battery Life:

~1yr 

Certification:

2yr NIST ISO17025 Certificate Included

Summary of Contents for SmartLOG 2021

Page 1: ...twork No cloud No subscription fees Enterprise Network Login for Security Time automatically synchronized from internert Reports in PDF format No PC Software needed Second Temperature Sensor optional...

Page 2: ...th 8 Lowest Tempeature Since Last Reset 9 Highest Temperature Since Last Reset 10 Alarm Set Points Min Max 11 Current Temperature P2 Tile Same as P1 1 Dashboard IP Address 2 Close Options Screen 3 Opt...

Page 3: ......

Page 4: ...d flashes two times then release button 5 Note the four digit serial number on the backside of the sensor s and note intended location for Fridge Freezer other E g Fridge SN 0010 6 Place wireless sens...

Page 5: ...being sent out When alerts are complete the screen will switch back to the main screen automatically showing DDL Tile blue and LOGGING OFF replaced wih Logger Name BUZZER Button When set to OFF will...

Page 6: ...RESET MIN MAX Button Sets memory recorded Min and Max values to current tempeature time and date and removes alert and alarm colors from display if excursion was triggered 1 Tap RESET MIN MAX button...

Page 7: ...batteries back in Push on button till you hear beep if no response data logger may be faulty Incorrect time Turn off logger Start logger will update from online time server NOTE DDL DOES NOT adjust fo...

Page 8: ...gging interval WiFi IP addr Mac addr Siginal strength CPU temperature Ram used and free Firmware version running Support Links to online Manual QuickStart video Instructional video REPORT PAGE Month s...

Page 9: ...rd Outgoing SMTP Dropdown list of popular email provider servers Selecting a server name will auto populate the Port Custom server address port can be entered If Alerts and Status and message are not...

Page 10: ...ccount click icon in upper right coner then click View Account On My Account page click Security info On My Sign Ins page click Add method Select App password click Add Enter password Name e g SmartLO...

Page 11: ...rview video Quick Start video Gmail Setup video SmartLOG Kit consist of 1 x SmartLOG Digital Data Logger 1 x Wireless Transmitter 1 x 5V 1A Power Adapter 1 x USB Cable 2 x AAA for transmitter 3 x AA f...

Page 12: ...Direct Sequence Spread spectrum Warranty With respect to the goods unless expressly agreed otherwise in writing Thermco Products Inc provides warranty for material and manufacturing defects for a peri...

Reviews: