11
2.
Press the TARE key repeatedly until the display reads "unlT"
3.
Press the COUNT/ENTER key, the display reads "G yes " which means g
(grams) is available for use.
4.
Press COUNT/ENTER to confirm. To disable g as the unit, press TARE key, the
display reads " G no " , press Count/Enter key to confirm.
5.
Follow the steps above to enable or disable Oz, Ct or dwt as the unit of
measure.
Restoring the Factory Default Setup
1.
Press and hold PRINT key and also press the % PERCENT key, the balance reads
"PrlnT"
2.
Press TARE key repeatedly until it reads "InlTlA"
3.
Press COUNT/ENTER key to confirm, the display reads "BUSY" and then returns
to weighing mode. Factory default setup is finished.