Compact Balances 

W3300 Series 

Operating Manual 

 

 

 

 

Summary of Contents for W3300 Series

Page 1: ...1 Compact Balances W3300 Series Operating Manual ...

Page 2: ...ements 5 2 3 Setting Up Your Balance 5 3 Operating your balance 6 3 1 Display and Control Panel 6 3 2 Basic Weighing 7 3 3 Count Weighing 8 3 4 Percent Deviation 8 4 Calibration 9 5 Print Setups 9 6 Communication with a Computer 12 6 1 RS232 Interface 12 7 Troubleshooting 13 8 Care and Maintenance 14 8 1 Cleaning 14 ...

Page 3: ...alance Multiple weighing modes 1 Standard weighing 2 Piece counting 3 Percentage weighing Choice of weighing units grams carats or ounces RS 232 interface for connecting peripheral device such as a computer or a printer Alarm to indicate malfunction Easy in lab calibration 2 INSTALLATION 2 1 UNPACKING CAUTION Always handle your Accuris balance with care The internal electromechanical components ha...

Page 4: ...d 500 only 1 5 AC power adapter 1 6 Operating Manual 1 Model W3300 120 W3300 300 W3300 500 W3300 1200 W3300 5K W3300 10K Capacity 120g 300g 500g 1200g 5000g 10 000g Calibration External Internal Quick Cal Readability 0 01g 0 1g Repeatability S D 0 02g 0 2g Linearity 0 02g 0 2g Stabilization time Approx 2 sec Operating temperature 10 30 Pan size 11 5cm 4 5 in 16cm 6 3i n 18cm 7in 18cm 7in Draft Shi...

Page 5: ...air conditioner o Near viberating rotating or reciprocating equipment o Near magnetic field or equipment that generates a magnetic field o On an unstable surface o In areas where there are corrosive vapors 2 4 SETTING UP YOUR BALANCE CAUTION Always allow the balance to warm up in its ambient environment for 2 hours prior to use 1 If the balance is moved from a cold environment to a warmer environm...

Page 6: ...p procedure the display will light up and the display will show 0 00 9 It is recommended to wait 30 minutes to allow the balance to equilibrate to the room s temperature before use 10 Note that the back light on the display will remain on when the balance is in use If the balance is not in use and there is not a sample on the weigh pan the back light will turn off as an energy saving feature The W...

Page 7: ... means 1 minute Press TARE repeatedly to toggle through the choices of the delay time Press ENTER to choose the value Now the backlight timer is set to the new value Display Information OK Weight reading shown is stable g Weight reading is shown in grams Oz Weight reading is shown in ounces Ct Weight reading is shown in carats Reading shown is given in as a percent weight see Percent Deviation Mod...

Page 8: ...d place the quantity of pieces onto the balance 2 Press the METRIC button and the display will indicate the chosen reference number of pieces 3 Remove the pieces from the weigh pan and the display should return to 0 4 Place a quantity of pieces onto the pan to count and the balance will calculate and display the number of the pieces 3 4 Percent Deviation This mode is used to calculate the percenta...

Page 9: ...icate that balance is ready for calibration 4 Place the appropriate calibration weight onto the weighing pan and wait for OK to display 5 The calibrated value for example 100 00 g will be displayed when the calibration is complete 6 Remove the weight and the display should return to 0 00 7 The balance is calibrated 5 Print Setups There are three print types STABLE Print print a stable reading when...

Page 10: ...ng the Baud Rate 1 Press and hold PRINT key and also press the PERCENT key the balance reads PrlnT 2 Press TARE key the balance reads bAud 3 Press COUNT ENTER key the balance reads 300 4 Press the TARE key repeatedly the balance display the other buad rates Choose the proper rate and press COUNT ENTER to confirm 5 The balance reads ParITy and begins parity check 6 Press the COUNT ENTER key the bal...

Page 11: ...he display reads G no press Count Enter key to confirm 5 Follow the steps above to enable or disable Oz Ct or dwt as the unit of measure Restoring the Factory Default Setup 1 Press and hold PRINT key and also press the PERCENT key the balance reads PrlnT 2 Press TARE key repeatedly until it reads InlTlA 3 Press COUNT ENTER key to confirm the display reads BUSY and then returns to weighing mode Fac...

Page 12: ...n below A B C D E F G H I J K L M A signs field usually no display as a space when it is a positve number is displayed when it is a negative number B G number and decimal field spaces are used when there are less than six digits H I spaces field J unit field it describes the units of the number being transmitted Your balance will transmit G for grams O for ounces C for carats K stable character it...

Page 13: ... 15 meters the load impedance of the device connected should be between 3000 and 7000 ohms with no more than 2500 pf shunt capacitance TROUBLE SHOOTING Problem Possible Cause Solution No Display No power AC DC Power Adapter not functioning or incorrect specifications Plug in the AC DC adapter Replace the adapter Contact the Service Department Displayed weight value is unstable Unsuitable environme...

Page 14: ...rument that must be handled carefully and properly maintained Do not use sharp objects such as a pen or pencil on the keypad Do not let objects fall on the weighing pan otherwise the weighing system can be damaged Do not expose the balance to high temperatures or dust Do not disassemble the balance without proper training or instructions It is recommended to cover the balance when not in use Keep ...

Page 15: ...15 ...

Page 16: ...16 Copyright 2015 Benchmark Scientific PO Box 709 Edison NJ 08818 USA www benchmarkscientific com www accuris usa com ...