11
2. Load a weight that is equal to or greater than the pre-set calibration weight on the plate; the
balance will recognize a weight that is equal to or greater than the calibration weight as valid
provided that it is a whole weight with respect to the most significant figure of the calibration
weight.
E.G.:
if the calibration weight is 200 g, it will be possible to calibrate the balance with values
that go from 200g, 300g, 400g up to the upper capacity limit of the balance.
The word
“LOAD”
on the display will stop flashing; once the calibration has been carried out,
the value of the weight used will be displayed.
3. Remove the calibration weight; the balance is ready for weighing operations.
NOTE: an error message will be displayed if there is some interference during the calibration process. To
interrupt the calibration process, press the ON/OFF button while the calibration weight indication flashes.
7.2 Balances with internal calibration
7.2.1 Internal calibration
In these models the calibration is carried out through an internal automatic system:
1. Press the
CAL
button with the weigh pan empty.
The display will show the message
“CAL”
and the balance’s calibration will be carried out
automatically.
200.0000
g
01-06-10 13:15:24
-0-
Load
01-06-10 13:15:24
-0-
Summary of Contents for W3101A-120
Page 60: ...60 TEST Weight 18 71 g Signature G L P parameters Weight data...