Do not attempt to open or repair the Mini Power Supply. 

Contact your distributor or Benchmark Scientific for service.   

 

3. Setup and Operation 

 

The power supply should be set up on a level and stable surface, 

such as a laboratory counter.    Choose a dry location that is not near 

a sink or running water. 

     
Set the On/Off switch to the middle “Off” position, connect the 
included power cord to the power inlet on the back of the instrument, 

and the other end of the cord to an appropriate electrical outlet.   

Check the rating label on the back of the power supply and make sure 

that the electrical supply from the outlet matches the requirements. 

The Mini Power Supply can accept input voltage from 100VAC to 

240VAC.   

 
With the power switch in the “Off” position, connect the 
electrophoresis gel box using electrical leads to the appropriate 

“+” 

(red) and 

“–“ (black) output jacks on the front of the power supply. 

 

CAUTION:

    Only use insulated electrical leads that are designed for 

electrophoresis applications with shielded banana connectors. 

 

CAUTION:

    Only use electrophoresis gel boxes that have 

appropriate safety covers and safety features to prevent electric 

shock to the user.   

 

CAUTION:

   

Always make sure that the gel box used has the 

appropriate buffer.    An incorrect concentration of buffer can cause 

excess current which will cause the fuse in the unit to blow. 

Summary of Contents for myVolt Mini

Page 1: ...myVoltTM Mini Power Supply Operating Manual ...

Page 2: ...Table of Contents 1 Introduction 2 Safety Warnings 3 Set up and Operation 4 Maintenance and Cleaning 5 Trouble Shooting 6 Specifications ...

Page 3: ...des product safety information and operating instructions Before using the Mini Power Supply please read this manual in its entirety and be sure to fully understand the features and methods for proper operation Keep this manual for future reference Please check the product carefully when first opening the box and if there are any parts missing or there is any damage please contact your distributor...

Page 4: ... it is operating to allow for sufficient air flow and cooling Always turn off the power after operation When not in use detach the power supply and store it in an area that is free from moisture and dust To avoid electrical shock do not use this product with wet hands Always use the Mini Power Supply in an environment with low humidity and low dust also keep it away from water direct sunlight stro...

Page 5: ...hat the electrical supply from the outlet matches the requirements The Mini Power Supply can accept input voltage from 100VAC to 240VAC With the power switch in the Off position connect the electrophoresis gel box using electrical leads to the appropriate red and black output jacks on the front of the power supply CAUTION Only use insulated electrical leads that are designed for electrophoresis ap...

Page 6: ...l maintenance Always disconnect the power cord and electrical leads for cleaning When cleaning the surfaces of the instrument use a cloth damp with water There are no user serviceable parts inside Do not open the casing of the Mini Power Supply 5 Troubleshooting Problem Cause Solution 1 No indicator light Power not connected Connect the power supply Fuse damaged Change a fuse Switch broken Contact...

Page 7: ...turing defects Model Specification Mini Power Supply E2100 Operating Temp range 4 C 40 C Input voltage AC100 240V 50 60Hz Output voltage DC 100 or 200 V Output error max DC 100V 4 5 DC 200V 7 2 Output power max 40W Ports 2 pairs Relative humidity 70 Dimension s mm 160x115x60 Weight kg 1 ...

Page 8: ... Copyright 2015 Benchmark Scientific PO Box 709 Edison NJ 08818 USA www benchmarkscientific com www accuris usa com Ph 908 769 5555 email info benchmarkscientific com ...

Reviews: