Troubleshooting {continued)
Problem
Causes
Resolutions
Error E1
Lid opened during run or
Seat lid properly.
Check
magnet switch problem.
magnet on lid ..
Lid not closed upon start or
Seat lid properly.
Check
Error E3
magnet on lid.
magnetic switch problem.
Check proper buffer level in
Electric current not flowing
the tank. Check buffer
concentration (1X or 0.5X
Error E5
normally. Buffer missing
,
or
concentration too high.
TAE or TSE is standard).
Check electrodes and
connections.
Buffer concentration too high,
Error E7
current too high
,
power over
Check buffer concentration.
50W, or internal electronic
Replace electronic board.
problem