OFS: Offset for % saturation measurement
NOTE:
this sub group appears in % saturation measurement mode only.
Use the offset adjustment to act as an offset at values between 0 and 100%. It is useful
to match against end-user standards (i.e. a test kit value).
8.4
From measurement mode:
1.
Press the Mode key
to select %
saturation mode.
2.
Press the Set up key
to enter Set Up
mode.
3.
Press the
▲
and
▼
keys
to scroll
through subgroups until you view
parameter “OFS” in the upper display.
See figure
4.
Press the ENTER key.
The upper
display shows the current measurement
in % saturation and the lower shows
OFS.
See figure
5.
Press the
▲
and
▼
keys
to offset the
% saturation measurement.
6.
Press the ENTER key
to confirm
selection and to move back to subgroup
“OFS”. If you want to return to measure-
ment mode, press CAL/MEAS.