background image

AM1860/1850 Standard Platinum Resistance Thermometer

User’s Guide

1/10

Table of Contents

Before you start ---- Warnings & Cautions ........................................................................ 2
1

Introduction................................................................................................................. 3

1.1

Main Application ................................................................................................ 3

1.2

Main Features...................................................................................................... 3

1.3

Calibrations ......................................................................................................... 3

2

Specifications .............................................................................................................. 4

2.1

Specifications ...................................................................................................... 4

3

General Operations .................................................................................................... 5

3.1

Connecting to the readout device........................................................................ 5

3.2

Drive Current ...................................................................................................... 5

3.3

Stability of Readings........................................................................................... 5

3.4

Immersion Requirements .................................................................................... 5

3.5

Thermal EMF...................................................................................................... 6

3.6

Over heating........................................................................................................ 6

4

Care and Handling Guidelines ................................................................................... 7

5

Troubleshooting .......................................................................................................... 8

5.1

Troubleshooting .................................................................................................. 8

5.2

Problem Causes and Solutions............................................................................ 8

6

Limited Warranty & Limitation of Liability................................................................ 9

Summary of Contents for AM1850

Page 1: ...AM1860 1850 Standard Platinum Resistance Thermometer User s Guide...

Page 2: ...3 Calibrations 3 2 Specifications 4 2 1 Specifications 4 3 General Operations 5 3 1 Connecting to the readout device 5 3 2 Drive Current 5 3 3 Stability of Readings 5 3 4 Immersion Requirements 5 3 5...

Page 3: ...ous live component 5 Follow all other safety guidelines listed in this user s guide Cautions Follow these guidelines to avoid possible damage to the instrument 1 Avoid mechanical shocks DO NOT drop or...

Page 4: ...ure temperature or change of temperature It has wide applications for dry wells or temperature baths 1 2 Main Features Extreme low drift rate Metal sheathed Great reference thermometer for dry block c...

Page 5: ...C 0 003 C 100 hours at 670 C R 0 01 C 0 01 C year Repeatability 0 0015 C Thermal Shock 0 0015 C after 10 times thermal cycles from minimum to maximum temperatures Self heating 0 0015 C at 1 mA current...

Page 6: ...wires can be distinguished by insulation colors Lead wire pairs attached to each end of the sensor are identified by red black and white blue insulation 3 2 Drive Current AccuMac recommends 1mA as dri...

Page 7: ...submerge PRT handle when taking measurement 3 5 Thermal EMF Each AccuMac PRT has gone through an annealing process and stability test to minimize the thermal EMF which is caused by either impurities...

Page 8: ...possible b When shipping the SPRT use protective box and other protective packaging materials to minimize mechanical shocks as much as possible c When using dry blocks make sure the well diameter is a...

Page 9: ...ir service Be sure to have the model number and serial number of your SPRT available 5 2 Problem Causes and Solutions a R0 Rtpw becomes higher significantly This is likely caused by mechanical shocks...

Page 10: ...e warranty period begins on the date of the shipment Parts product repairs and services are warranted for 90 days The warranty extends only to the original buyer or end user customer of an AccuMac aut...

Reviews: