AM1710 Secondary Reference PRT User’s Guide

 

 

 

 

Rev. 201803V1

 

2/10

Before you start ---- Warnings & Cautions

 

 

 

 

 

Warnings: 

Follow these guidelines to avoid personal injury:

 

1.

 

Only use this instrument in the manufacture specified 
temperature range. 

2.

 

The handle of this instrument can become hot when it is 
used to measure high temperatures for extended periods of 
time. 

3.

 

DO NOT submerge PRT handle when taking 
measurement. 

4.

 

DO NOT use this instrument to measure the temperature 
of any hazardous live component. 

5.

 

Follow all other safety guidelines listed in this user’s 
guide. 
 

 

Cautions: 

Follow these guidelines to avoid possible damage 

to the instrument:

 

1.

 

Avoid mechanical shocks. DO NOT drop or slam the 
probe in any way. This will cause damage to the probe 
internally and affect its calibration and accuracy. 

2.

 

Read Section entitled “Care and Handling Guidelines” 
before removing the PRT from the shipping box. Incorrect 
handling can damage the PRT and void the warranty. 

3.

 

Keep the shipping container in case it is necessary to ship 
the PRT. Incorrect packaging of the PRT for shipment can 
cause irreparable damage. 

4.

 

Calibration Equipment should only be used by Trained 
Personnel. 
 

 

 

 
 

Summary of Contents for AM1710

Page 1: ...Rev 201803V1 AM1710 Secondary Reference Platinum Resistance Thermometer User s Guide Please download the latest version at www accumac com...

Page 2: ...librations 3 2 Specifications 4 2 1 Specifications 4 3 General Operations 5 3 1 Connecting to the readout device 5 3 2 Drive Current 5 3 3 Stability of Readings 5 3 4 Immersion Requirements 5 3 5 Ther...

Page 3: ...live component 5 Follow all other safety guidelines listed in this user s guide Cautions Follow these guidelines to avoid possible damage to the instrument 1 Avoid mechanical shocks DO NOT drop or sla...

Page 4: ...temperature or change of temperature It has wide applications for dry wells or temperature baths 1 2 Main Features High accuracy 0 015 C at 0 01 C Temperature range 60 C to 160 C Durable and shock re...

Page 5: ...9 seconds for 63 response to step change in water moving at 3 feet per second Measurement Current 0 5 mA or 1 mA Sensor Length 32 mm Sensor Location 5 mm from tip Insulation Resistance 1000 M at room...

Page 6: ...hed to each end of the sensor are identified by red black and white blue insulation 3 2 Drive Current AccuMac recommends 1mA as drive current to ensure the best measurement 3 3 Stability of Readings T...

Page 7: ...cuMac PRT has gone through an annealing process and stability test to minimize the thermal EMF which is caused by either impurities of sensing element or temperature differentials at lead wires connec...

Page 8: ...ssible b When shipping the PRT use protective box and other protective packaging materials to minimize mechanical shocks as much as possible c When using dry blocks make sure the well diameter is appr...

Page 9: ...service Be sure to have the model number and serial number of your PRT available 5 2 Problem Causes and Solutions a R0 Rtpw becomes higher significantly This is likely caused by mechanical shocks The...

Page 10: ...ty period begins on the date of the shipment Parts product repairs and services are warranted for 90 days The warranty extends only to the original buyer or end user customer of an AccuMac authorized...

Reviews: