40-6910 Service Manual
3.0 Trouble Shooting Guide
No.
Symptom
Cause
Corrective Action
1
The horizontal micro-motion knob
doesn’t work
The horizontal locking mechanism is
damaged or defective
Replace the horizontal locking
mechanism
2 The focus knob doesn’t work
The upper gear on the focus tube has a lot
of wear
Replace the focus tube
3 Fail to level the bubble
The bubble is broken
Replace the bubble
4 The cross hair is obscure
The position of the eyepiece is not correct Adjust the position of the eyepiece
5 There is something on the reticle
The reticle is dirty
Clean the reticle
6
Bubble is not centered in any
direction
The bubble level screw is loose
Recalibrate according to the above
method