Prepared By: Tim Wojo
Revision - Date: 0 – 03232009
40-6537 Service Manual
SLOPE
SLOPE
SLOPE
SLOPE
1. Enter self-calibration mode
Power off the instrument and direct the X-axis towards wall.
Press the key and key simultaneously. While continuing to hold the key, remove finger from
the key. While continuing to hold the key, observe the main display’s LED’s flash 3 time. Once this
happens, remove finger from the key and observe that the rotating head of the laser continuously rotates. The
laser is now in calibration mode.
2. X-axis calibration
Press the key one time and verify that the X-direction LED on the keypad of the 40-6537 flashes,
indicating that it has entered the self-calibration mode on the X-direction.
Press the key to adjust the laser beam up or down to coincide with the reference line.
3. Y-axis calibration
Rotate the 40-6537 90° so that the Y-axis is now facing the wall.
Press the key one time and verify that the Y-direction LED on the keypad of the 40-6537 flashes,
indicating that it has entered the self-calibration mode on the Y-direction.
Press the key to adjust the laser beam up or down to coincide with the reference line.
4. Self-calibration confirmation
Press the key after finishing self calibration on both the X and Y axis and verify that the Self Calibration LED
extinguishes. The instrument will exit the self-calibration mode automatically after finishing self-calibration.
Power off the unit to store the calibration values into the 40-6537’s memory.
Repower the 40-6537 and confirm the units accuracy. Repeat procedure as required.