background image

13

To configure the balance/scale; i.e., adapt the balance/scale to individual requirements.

Step

Key (or instruction)

Display

1. Switch off the balance/scale 

(ON/OFF)

2. Switch on the balance/scale and 

(ON/OFF)

while all segments are displayed

(ZERO) briefly

Navigation in the Setup Menu

Key

Press briefly

Press and hold

(ENTER)

Menu level: move to the right (cyclical)

Confirm setting

(F)

Menu item: Scrolling 

(ZERO)

Menu level: Move to the left

Save settings and exit Setup

Parameter Settings (Overview)

ο

Factory setting

√ 

User-defined setting

Setup

1

Weighing

1.1. Adapt filter

1.1.1

Very stable conditions

menu

1.1.2

ο

Stable conditions

1.1.3

Unstable conditions

1.1.4

Very unstable conditions

1.2. Application filter

1.2.1

ο

Final readout

1.2.2

Filling

1.3. Stability range

1.3.1

1/4 digit

1.3.2

1/2 digit

1.3.3

1 digit

1.3.4

ο

2 digits

1.3.5

4 digits

1.5. Calib./adjust./linearization:

1.5.1

ο

Calibration/adjustment

Function of the (CAL) key

1.5.2

Linearization: for service personnel only

1.5.3

Key blocked

1.6. Auto zero

1.6.1

ο

On

1.6.2

Off

1.7. 1st weight unit, or 2nd unit in 

1.7.1

User-defined unit; see 

Toggle Weight Units app.

to 1.7.23

“Toggling between Weight Units”

5. and 6. 

Only relevant with built-in data interface: see corresponding interface description 

8. Additional

8.1. Block key functions

8.1.1

All keys blocked except for (ON/OFF)

functions

and (ZERO)

8.1.2

ο

All keys unblocked

8.2. Automatic shut-off 

8.2.1

ο

After 2 minutes

8.2.2

After 5 minutes

8.2.3

After 10 minutes

9. Reset menu

9.1. Factory settings

9.1.1

Restore

9.1.2

ο

Do not restore

Configuration (Setup Menu)

Summary of Contents for ACCULAB VICON VIC-123

Page 1: ...de d emploi Istruzioni per l uso Manual de instrucciones ACCULAB VICON Electronic Precision Scales Balances Elektronische Pr zisionswaagen Balances lectroniques d analyse et de pr cision Bilance elett...

Page 2: ...ies Options 16 C Marking 2 Contents Warnings and Safety Precautions Safety Information To prevent damage to the equip ment please read these operating instructions carefully before using your balance...

Page 3: ...adjustable feet Storage Do not stack more than 3 balances on top of one another at a time Installation Choose a location that is not subject to the following nega tive influences Heat heater or direct...

Page 4: ...lance scale on its side Open the battery compartment remove the compartment cover Install the battery in the compart ment Make sure the polarity is correct Close the battery compartment slide the cove...

Page 5: ...off automatically unless a key is pressed Description of the Keys ON OFF On off key switches the balance scale on and off or switches it to the standby mode Battery operation on turns backlight on of...

Page 6: ...e on balance scale 7 Toggle weight unit F Menu code Unit Conversion factor Display 1 user Grams 1 00000000000 o 2 grams factory setting Grams 1 00000000000 g 4 carat Carats 5 00000000000 o 5 Pound Pou...

Page 7: ...ale 5 Zero the balance scale ZERO 6 Place reference sample quantity 20 on the balance scale 7 Select reference sample quantity in increments of 1 1 2 3 99 or F repeatedly briefly or in increments of 1...

Page 8: ...mpty container on the balance scale 5 Tare the balance ZERO 6 Place the reference weight for 100 on the balance scale 7 If desired change the number of decimal places displayed 100 0 100 00 or 100 fac...

Page 9: ...n the display application is active 4 If necessary zero the balance scale ZERO 5 Place oversized sample on balance scale 6 Start application program F Symbol flashes on the display the weight value is...

Page 10: ...yed ENTER steadily symbol flashes 7 Remove sample from balance scale 8 Place the next sample on the balance scale in this example 575 g 9 Store value in memory Totalized stored weight ENTER is display...

Page 11: ...ance scale ZERO 5 Start application program F 6 Confirm the display airval ENTER 7 Determine weight of the sample in air place sample on the balance scale 8 Store weight in air value ENTER 9 Remove sa...

Page 12: ...different calibration weight value F repeatedly 5 Confirm calibration weight value and ENTER start calibration span adjustment After the zero point is stored the required calibration weight flashes on...

Page 13: ...ons 1 2 Application filter 1 2 1 Final readout 1 2 2 Filling 1 3 Stability range 1 3 1 1 4 digit 1 3 2 1 2 digit 1 3 3 1 digit 1 3 4 2 digits 1 3 5 4 digits 1 5 Calib adjust linearization 1 5 1 Calibr...

Page 14: ...play calErr Calibration parameter not met e g Calibrate only when zero is displayed Balance scale not zeroed Press ZERO to tare the balance scale Balance scale is loaded Unload the balance scale appEr...

Page 15: ...of operation w 9V battery Alkaline approx h 11 11 14 Rechargeable fully ch NiMH avg h 2 5 2 5 4 Model VIC 5101 VIC 3101 VIC 1501 VIC 711 VIC 511 VIC 10KG VIC 6KG VIC 4KG Weighing capacity g 5100 3100...

Page 16: ...request Accessories Options C Marking The balance scale complies with the following EC Directives and European Standards Council Directive 89 336 EEC Electromagnetic compatibility EMC Applicable Europ...

Reviews: