Accud CF1250 Operation Manual Download Page 2

 

 

 

8

 

 

 

 

Fig. 4-2 

"Down"

  ("Up" or "Down" 

  the same below) 

can select a item, press "OK"

  to enter the selected 

item. 

In the menu settings, press to exit or return to the 

previous menu. You can also press 

  longer to exit 

setup and return to the measurement mode. 
 

4.1 Measurement Settings 

Select Measure to enter the 
measurement parameter 
settings. 
Measurement parameters 
include: measurement mode 
(Mode), upper and lower limit 
settings (Tolerance) and 
calibration (Calibration). 
 
4.1.1 Measuring mode 

a) Press 

to select Mode, press

b) Press 

to select Single (Single Mode) or Continue 

(Continuous Mode); 

c) Press 

  to save and return to the measurement 

mode; 

d) Press 

  to return to the measurement mode and not 

save. (Fig. 4-2) 

9

 

 

 

 

 

Fig. 4-3 

4.1.2 Upper and lower limit settings 

If the limits are set up and turn on, when the measuring 
value exceeds the limit, the 
instrument will audible alarm. 
(Fig. 4-3) 

a) Press

  to select 

Tolerance (upper and lower 

limit settings), press

  to 

enter; 

b) Press 

  set to "On" or 

"Off"; 

c) Press

  to select Upper 

limit or Lower limit, press 

to enter to input the upper 

and lower limits; 

d) Press 

to increase or 

decrease the setting, press 

  to move to the next digit; 

Note: The upper limit must be greater than the lower 
limit, otherwise the setting is invalid. 
 
4.1.3 Calibration 

(Detailed calibration procedure See 5.1) 

 

4.2 System Settings 

Select System, can enter the system settings. 
System setting parameters include: unit (Unit), backlight 

10

 

 

 

 

Fig. 4-4 

 

 

Fig. 4-5 

and key tone (Sound). 
 

4.2.1 Unit 

a) Press

  select Unit, 

press

b) Press 

to select "μm" or 

"mils"; 

c) Press 

to save the settings 

and return to measurement 

mode;(Fig. 4-4) 
 

4.2.2 Backlight 

Set backlight to "Off" will be no backlight, set to "On", the 
backlight will be opened 
constantly. 

NOTE: Setting the backlight to "On" is 
not recommended for saving battery 
power. (Fig. 4-5) 

a) Press 

to select Backlight, 

press

b) Press 

to select "Off" or 

"On"; 

c) Press 

to save the settings 

and return to measurement 

mode; 
 
 

11

 

 

 

 

Fig. 4-6 

4.2.3 Key Tone 

(Fig. 4-6) 

a) Press 

to select Sound 

(touch-tone), press

  ; 

b) Press 

  to select "On" 

(Turn on key tones) or "Off" 
(turn off the key tone); 

c) Press

  to save the 

settings and return to 

measurement mode. 

 

5    Calibration and Maintenance 

5.1 Calibration 

For ensuring measurement accuracy, you may need 
calibrate to correct the error caused by wear and tear or 
changes in environment etc.. 

Press 

to enter the menu and select Measure, press 

to select Calibration, you can enter the instrument 

calibration interface. 
Three calibration methods are provided:    Zero calibration 
(Zero), One-point calibration (1st) and Two-point 
calibration (2nd). 

NOTE:    Before One-point calibration zero calibration must be 
carried out.    Before the Two-point calibration, the 
Zero-calibration and One-point calibration must be carried out. 

 

12

 

 

 

 

Fig 5-1 

5.1.1 Zero Calibration 

Zero calibration may be in two ways: Zeroing and Menu 
zero calibration. 

1) Zeroing 

a) In single mode, measure on the substrate directly, 
display the results "xx μm"; 

b) Press 

  to clear display as "0.0 μm". 

NOTE: 
 i) If "xx μm" is greater than 80μm, you cannot zero it directly, 
Menu zero calibration is needed. 
ii) Repeating the above a), b) as a few times may increase the 
accuracy of the calibration. 

 
2) Menu Zero Calibration 

(Fig. 

5-1)

 

a) Enter Menu and select 

Calibration, then select Zero (zero 

calibration), press

  , wait for 

showing Acting; 
b) Measure on the substrate 
directly, measuring 3 to 5 times or 
more can ensure accuracy; 

c) Press

  to save the 

calibration values and remain in the calibration interface. 

Then press 

  can finish the Zero calibration and return; 

d) If press 

  only, will be not save the calibration and 

return

13

 

 

 

 

 

Fig. 5-2 

5.1.2 One-point calibration

 

Before One-point calibration 
perform, must first do 
Zero-calibration.(Fig. 5-2) 
a) Press

to select the 1st 

(One-point calibration), "0000.0" 
will be shown; 
b) Press

to enter the standard 

value of foil; 

c) Press 

to adjust the value, 

press 

  to move to the next 

digit, Input the standard thickness 

value of the foil, (press 

  to 

return to the initial setting value). 

Then, press 

  and waiting for 

showing Acting; 
d) Measuring on the standard thickness foil for 3 to 5 
times; 

e) Press

  to save the calibration values and remain in 

the calibration interface, you can proceed to the next one 

calibration; 

f) Or press 

  to not save the calibration and return. 

 
5.1.3 Two-point calibration 

Before Two-point calibration, first the Zero-point 
calibration and the Two-point calibration must be done. 
And thickness of the standard foil must be greater than in 

14

 

 

 

 

 

Fig 5-3 

the One-point calibration.(Figure 5-3) 
a) Press

to select the 2nd (Two-point calibration), 

"0000.0" will be shown; 
b) Press

to enter the standard value of foil; 

c) Press 

to adjust the value, press 

  to move to 

the next digit, Input the standard thickness value of the 

foil, (press 

  to return to the initial setting value). Then, 

press 

  and waiting for showing Acting; 

d) Measuring on the standard 

thickness foil for 3 to 5 times; 

e) Press

  to save the 

calibration values and remain in 

the calibration interface, you can 

proceed to the next one 

calibration; 

f) Or press 

  to not save the 

calibration and return. 

 

5.2 Maintenance 

1) Avoid dirt, dust, humidity, 
corrosive substances of chemical 
environments. Prevent the 
instrument has been dropped! 
2) After use, place the instrument back into the protective 
device. 

15

 

 

3) Avoid strong sunlight and temperature fluctuations, 
which would adversely affect the measurement results. 
4) The instrument enclosure resistant to most chemical 
cleaners, use a soft cloth to clean the instrument. 
5) Only keep it clean probe to measure the accurate results, 
so the probe regularly inspect and clean them paint, iron 
and other residues. 
6) Standard thickness foil is very important in the calibration. 
So if the standard pieces appear worn, bent or damaged, 
contact your dealer to replace. 
7) If the gauge is not working, please do not disassemble it. 
Should send it to the professionals or back to the factory for 
repair. 

 

                                                                                                                 

Reviews: