5). Wipe the piston, the 0-ring and the tip holder
with ethanol and a lint-free cloth.
Note: Same models have a fixed 0-ring lo
cated inside the tip holder. In this case, the 0-
ring can not be accessed for maintenance.
6). Before replacing the tip holder, it is recom
mended to grease the piston slightly using the
grease provided.
Note: Excessive use of grease may jam the
piston.
7). After reassembling, use the pipette without
liquid several times to make sure that the grease
is spread evenly.
8). Check the performance and recalibrate the
pipette.
10. TROUBLE SHOOTING
A quick inspection of the pipette referring to the
below questions may help you isolate the problem.
- ls the push button rod straight?
- ls the push button free of chemical damage?
- Are the volumeter digits clear and easy to read?
- Can you set any volume in the useful range?
- Can you set the maximum volume?
- Does the volumeter operate smoothly without
hitching?
lf the answer to any of these questions is NO, the
pipette probably has been damaged as the result
of a mechanical shock or chemical damage, in
which case it should be returned to your supplier
9