EXI-600 INVERTED MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
13
5.2.5.
Stage
Tilt the transmitted illumination
pillar into its back position.
Align the holes in the stage with
the screw holes in the mainframe.
NOTE: The stage X-Y control arm
should be positioned to the right.
NOTE: 2 screw holes are located in
the front of the stage closest to the
observation tube, and 1 screw hole
centered in the back by the light pillar.
Use a 3mm hex screwdriver to
secure the stage with the supplied
3mm hex screws.
5.2.6.
Objectives
Install the objectives from low
magnification to high, beginning
with the lowest position number
on the nosepiece.
If installing different types of
objectives (e.g., brightfield and
phase contrast), install all of one
type before installing the next.
Remove a dust plug from the
nosepiece. Screw an objective
into the threaded opening.
Note the position number on the
nosepiece that corresponds to the
objective just installed. This will
be needed when programming the
LCD screen later.
Do not over tighten the objective.
Hand tight is sufficient.
Repeat until all objectives are
installed.
To minimize dust from getting on the optics below, leave dust plugs in place where objectives are not
installed.
3mm hex screws
Stage
Illumination pillar
Stage X-Y control arm
Dust plug
Objective