background image

 

EXI-310 INVERTED MICROSCOPE SERIES   

 

 

 
 

ACCU-SCOPE

®

     

                   

73 Mall Drive, 

Commack, NY  11725  •  631-864-1000  •  www.accu-scope.com 

7

 

ASSEMBLY DIAGRAM  

The diagram below shows how to assemble the various components. The numbers indicate the order of 
assembly. Use the 1.5mm and 2mm hex wrenches that are supplied with your microscope when 
required.  Be sure to keep these wrenches for changing out components or making adjustments. 
 
When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching 
any parts or touching glass surfaces. 

 

 

 
 

 

 

 

 

 

 

① Objectives 

② Phase Slider 

③ Stage Plate 

④ Eyepieces 

⑤ Power Cord

 

Summary of Contents for EXI-310 Series

Page 1: ...MANUAL 73 Mall Drive Commack NY 11725 631 864 1000 P 631 543 8900 F www accu scope com info accu scope com EXI 310 INVERTED MICROSCOPE SERIES...

Page 2: ......

Page 3: ...IAGRAMS 5 MICROSCOPE DIMENSIONS 6 ASSEMBLY DIAGRAM 7 DETAILED ASSEMBLY 8 9 OPERATION ILLUMINATION 10 INTERPUPILLARY DISTANCE 10 FOCUSING 10 ADJUSTABLE TENSION CONTROL 11 STAGE PLATES OPTIONAL 11 SELEC...

Page 4: ...component including eyepieces objectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More p...

Page 5: ...ipment of your microscope if needed Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew will form Carefully remove the microscope from the EPE f...

Page 6: ...ive Nosepiece Ring Coarse Fine Focus Adjustment Knob Base Eyepieces Phase Contrast Slider Condenser Stage Camera Photo Port Optional Fluorescence Light Path Access Panel NOTE If you did not purchase y...

Page 7: ...EXI 310 INVERTED MICROSCOPE SERIES ACCU SCOPE 73 Mall Drive Commack NY 11725 631 864 1000 www accu scope com 6 MICROSCOPE DIMENSIONS...

Page 8: ...rs indicate the order of assembly Use the 1 5mm and 2mm hex wrenches that are supplied with your microscope when required Be sure to keep these wrenches for changing out components or making adjustmen...

Page 9: ...epiece is at its lowest position 2 Remove the nosepiece cap closest to you and thread the lowest magnification objective onto the nosepiece opening then rotate the nosepiece clockwise and thread the o...

Page 10: ...OLTAGE CHECK Confirm that the input voltage indicated on the rear label of the microscope corresponds to your line voltage The use of a different input voltage than indicated will cause severe damage...

Page 11: ...bes coincide completely A complete circle should be seen in the viewing field when viewing the specimen slide An improper adjustment will cause operator fatigue and will disrupt the objective parfocal...

Page 12: ...ent ring Turn the tension adjustment ring clockwise to loosen or counterclockwise to tighten according to user preference Fig 8 Using the Stage Plates Optional Fig 9 10 NOTE for optimal viewing ensure...

Page 13: ...ition pushed all the way to the right Slide the light path selection slider to the left to the IN position This will send 20 of the light to the binocular viewing head and 80 to the top photo port for...

Page 14: ...specimen which is not dyed set the iris diaphragm lever to 75 of the N A of the objective in use Phase Contrast Observation Depending on the configuration ordered the EXI 310 can be used for phase co...

Page 15: ...Annulus Fig 15 16 The phase slider is pre aligned at our facilities If realignment is necessary follow these steps 1 Place a specimen on the stage and bring it into focus 2 Replace the eyepiece in the...

Page 16: ...unt of media or wellplate inconsistences This can reduce image contrast If this occurs repeat steps 1 5 for readjustment The centering procedure may have to be repeated in order to get the best possib...

Page 17: ...rast observation Dust caps for slider and push rod for filter position and selection Each position of the slider clicks securely into position due to click stop ball bearing positioning Alternate ligh...

Page 18: ...kwise to close it Observe your specimen through the eyepieces and open the field iris diaphragm gradually to find the image in the field of view If your image is not centered in the field of view adju...

Page 19: ...poor Details are indistinct Phase contrast effect cannot be obtained The objective is not correctly engaged in the light path Turn the nosepiece into the engaged position the aperture diaphragm is op...

Page 20: ...er or call ACCU SCOPE at 1 888 289 2228 for an authorized service center near you The light intensity is not enough Not use an designated lamp use an designated lamp The brightness adjustment knob is...

Page 21: ...ce to the original end user purchaser LED lamps are warranted for a period of two years from the date of original invoice to the original end user purchaser The mercury power supply is warranted for a...

Reviews: