ADJUSTMENT & OPERATION

 (continued)

 

 

 

Darfield stop in closed position 

 

 

Condenser in lower position 

 

Using the Darkfield Stop (optional) 

1.  To view specimens using the Darkfield Stop, rotate the stop into the 

closed position. 

2.  While viewing a specimen, adjust the iris diaphragm open or closed 

to optimize the image.    The specimen should appear mostly white 

on a dark background. 

3.  To view the specimen in brightfield mode, rotate the darkfield stop 

into the open position. 

 

 

NOTE: To move the darkfield stop in or out of the light path, you 

may need to lower the condenser.    To lower the condenser, 

grasp the metal ring 

 on the condenser and rotate 

counterclockwise (to the left).    After changing the position of 

the darkfield stop, raise the condenser back to the upper limit 

just below the stage. 

 

2. 

Darkfield stop in open position

 

 

Fig. 17 

 

Storing the Power Cord 

When the microscope is not in use, the power cord can be wrapped 
around the cord wrap on the back of the microscope, and the power 
charger can be plugged into the socket on the back of the microscope to 
avoid being lost. (See Fig. 17) 

 

Do not use strong force when the power cord is bent or twisted, 
otherwise it will be damaged.   

 

Use the power cord provided with your microscope by 
ACCU-SCOPE.    If it is lost or damaged, a replacement can be 
purchased from ACCU-SCOPE.    Always choose a power cord with 
the same specifications as the original.

 

 

Summary of Contents for EXC-100 Series

Page 1: ...73 Mall Drive Commack NY 11725 631 864 1000 www accu scope com info accu scope com MANUAL EXC 100 MICROSCOPE SERIES...

Page 2: ...G THE 100X OIL IMMERSION OBJECTIVE 11 A MOVE THE EYEPIECE TO EXAMINE THE AIR BUBBLE OPEN THE APERTURE DIAPHRAGM AND FIELD DIAPHRAGM FULLY AND OBSERVE THE EDGE OF THE OBJECTIVE FROM THE TUBE IT SEEMS R...

Page 3: ...t including eyepieces objectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persisten...

Page 4: ...icroscope from the Styrofoam container by its arm and base and place the microscope on a flat vibration free surface Operation Notice Fig 1 1 As the microscope is a high precision instrument always op...

Page 5: ...SERIES 73 Mall Drive Commack NY 11725 631 864 1000 www accu scope com 5 COMPONENTS DIAGRAM Objective Stage Illuminator Fine Focusing Knob Coarse Focusing Knob Focus Tension Adjustment Ring Frame Binoc...

Page 6: ...Y 11725 631 864 1000 www accu scope com 6 Eyepiece Trinocular Head Fine Focusing Knob Coarse Focusing Knob Frame Objective Stage Condenser Lock Screw Indicator Light Light Adjustment Knob Nosepiece Ve...

Page 7: ...ians at our New York facility prior to shipment Should you need to disassemble assemble your microscope in the future please follow the instructions outlined below When assembling the microscope make...

Page 8: ...objective nosepiece until it sounds ka da to make sure the desired objective is in the center of optical path Fig 4 Installing the Eyepiece s 1 Remove the cover from the eyepiece tube 2 Insert the ey...

Page 9: ...connected to ground ADJUSTMENT OPERATION Fig 7 Illumination 1 Turn on the power by rotating the light adjustment knob and adjust light intensity by rotating the knob until the illumination is comforta...

Page 10: ...focus ring located just below the right eyepiece until the image seen in the right eyepiece matches the left eyepiece The focus travel locking screw sets the maximum height that the stage can move wh...

Page 11: ...ted into the light path Fig 12 Using the 100x Oil Immersion Objective 1 Use the 4X objective to focus the specimen 2 Place a drop of oil on the specimen observed see Fig 12 3 Rotate the nosepiece coun...

Page 12: ...ssembling and Using the Trinocular Port and Camera trinocular model only 1 Loosen the lock screw of trinocular head and remove the dustcover See Fig 14 2 Remove the dustcovers from the camera adapter...

Page 13: ...and allow the clip to slowly close onto the slide and clamp the slide firmly see Fig 15 Do not let go of the lever as the clamp will close too fast This may result in slide breakage and flying glass...

Page 14: ...wer the condenser To lower the condenser grasp the metal ring on the condenser and rotate counterclockwise to the left After changing the position of the darkfield stop raise the condenser back to the...

Page 15: ...ion by swinging the objective correctly into the optical path Dirt or dust on the view field Dirt or dust on the lens eyepiece condenser objective collector lens or specimen Clean the lens Poor image...

Page 16: ...ating the intensity control dial and or iris diaphragm Open to the proper setting Position the condenser slightly lower than the upper limit MECHANICAL PROBLEMS Image will not focus with high power ob...

Page 17: ...e of the condenser is too low Raise the condenser base Slide does not move smoothly The slide is not placed correctly The movable specimen holder is not mounted properly Place the slide in the holder...

Page 18: ...Commack NY 11725 631 864 1000 www accu scope com 18 enough The light adjustment knob is set too low The light adjustment knob is not working correctly one Adjust the light intensity correctly Replace...

Page 19: ...nship for a period of five years from the date of invoice to the original end user purchaser The LED lamp is warranted for a period of one year from the date of invoice to the original end user purcha...

Reviews: