TROUBLESHOOTING GUIDE 

 

If a problem occurs during the course of use, please refer to the tables below                

before contacting your ACCU-SCOPE distributor. 

 

 

 

OPTICAL 

Problem 

Cause 

Corrective Measure 

Darkness at the periphery 

Revolving nosepiece not in  

Revolve the nosepiece to 
click-stop position by 

or uneven brightness in the  

click stop position 

swinging the objective 

field of view 

  

correctly into the optical path 

  

  

 

  

  

  

Dirt or dust on the viewfield 

Dirt or dust on the lens -  

Clean the lens 

  

eyepiece, condenser,  

  

  

objective, collector lens or 

  

  

specimen 

  

  

  

  

Poor image quality 

No coverglass attached to  

Attach a 0.17mm coverglass 

  

the slide 

  

  

  

  

  

Coverglass is too thick or thin 

Use a coverglass of the  

  

  

appropriate thickness 
(0.17mm) 

  

  

  

  

Slide may be upside down 

Turn slide over so the cover- 

  

  

glass faces up 

  

  

  

  

Immersion oil is on a dry  

Check the objectives, clean 

  

objective (especially the 40xR) 

if necessary 

  

  

  

  

No immersion oil used with 

Use immersion oil 

  

100xR objective 

  

  

  

  

  

Air bubbles in immersion oil 

Remove bubbles 

  

  

  

  

Condenser aperture is closed 

Open or close properly  

  

or open too much 

  

  

  

  

  

Condenser is positioned too 

Position the condenser at 

  

low 

the upper limit 