TROUBLESHOOTING GUIDE

 

  

PHASE CONTRAST MICROSCOPY

 

PROBLEM

 

CAUSE

 

CORRECTIVE MEASURE

 

Poor phase 

contrast image    

is obtained

 

The condenser phase annular 
diaphragm image and the objective 
phase plate do not match.

 

Adjust the phase annular 
diaphragm so that it matches with 
the objective phase plate. 

 

The condenser phase annular 
diaphragm and the objective phase 
code do not match. 

 

Insert the phase annular 
diaphragm with the same phase 
code as the objective into the 
optical path.

 

The field diaphragm image is not 
focused on the specimen surface.

 

Move the condenser up or down to 
focus the image on the specimen 
surface.

 

The phase difference of the 
specimen is too large.

 

Prepare the specimen using a 
different sealant and specimen 
thickness

 

The type of phase objective does 
not match the phase difference of 
the specimen.

 

Select a phase objective suitable 
for the specimen. 

 

 