Accu-Flo Meter Service MID COM 8000 Calibration Instructions Download Page 2

 

2

6. 

After setting the switches to the above values indicated, you now want to move 

 

switch # 7 (toggle switch) from the centre or net position to the full left or gross 

 position. 
 
7. 

With switch # 7 (toggle) in the gross or left position, you now want to fully 

 

depress switch # 8 (plunger reset). The register will now go through a reset and 

 

you will see the numbers 1 through 9 successively appear on your screen. 

 
8. 

After the register has reset, you now want to move switch # 7 “back” to the 

 

centre or “NET” position (

“VERY IMPORTANT”

 
9. 

We now want to manually reset the register by moving the # 7 toggle switch all 

 

the way to the right, this position is spring loaded and will return to the center or 

 

net position.  The register will now go through a reset and you will see the 

 

numbers 1 through 9 successively appear on your screen. 

 
 

SECTION TWO 

 
 
1. 

With the computer in the cab still set to “GASOLINE” 

 
2. 

Now we want to move switch # 6 to the number “9” position, and again 

 

manually reset the register by moving the # 7 toggle switch all the way to the 

 right. 
 
3. 

Now record the numbers that appear on the registers display; 

 
 

GASOLINE   

 

 

 

 
 _______________________ 
 
 
4. 

Now at the computer in the cab, press the line 2 key to change the product to 

 

“DISTILLATE”, and then check the registers display, and record the numbers 

 

that appear below; 

 
 

DISTILLATE   

 

 

 

 
 _______________________ 
 
 