Put the disconnect cover (L) on the cannula
housing.
Reconnect the infusion set
WARNING
Before reconnecting, ensure that a drop of insulin
appears at the tip of the connector needle.
Prime the infusion set until insulin comes out of
the connector needle.
Hold the self-adhesive (F) in place and remove
the disconnect cover (L) from the cannula housing
(G). Re-attach the connector (H) to the cannula
housing. An audible “click” ensures the connector
is locked.
WARNING
It is not necessary to give an additional bolus.