Patient’s blood glucose may be 
extremely high or a meter or test 
strip error has occurred.

• 

If the patient’s blood glucose 
result matches how he/she 
feel, contact the doctor 
immediately.

• 

If the patient’s test result does 
not match how he/she feel, 
repeat the blood glucose test.

• 

If the E-3 code still appears for 
patient’s blood glucose test, 
his/her blood glucose result 
may be extremely high and 
above the system’s reading 
range. Contact the doctor 
immediately.

Not enough blood or control 
solution was drawn into the test 
strip for measurement or was 
applied after the test had started. 
Discard the test strip and repeat 
the blood glucose or control test.

Blood or control solution was 
applied to the test strip before the 
fl ashing drop symbol appeared on 
the display. Discard the test strip 
and repeat the blood glucose or 
control test.

The meter is ready for you to 
insert a test strip.

Troubleshooting

Error messages

The test strip may be damaged 
or not properly inserted. 
Remove and reinsert the test 
strip and replace it if damaged.

The meter will not turn on or the 
display is blank.

• 

Battery is dead. Insert new battery.

• 

Display is damaged. Contact 
Roche.

• 

Meter is defective. Contact Roche.

• 

Extreme temperatures. Move the 
meter to a more temperate area.

The meter is in set-up mode, 
waiting for you to change or 
confi rm settings.

Battery power is low. Change the 
battery soon.

The meter is ready for a drop 
of blood or control solution.

Blood glucose may be lower than 
the measurement range of the 
system (0.6 mmol/L). Repeat test.

Blood glucose may be higher than 
the measurement range of the 
system (33.3 mmol/L). Repeat test.

An electronic error occurred, or 
in rare cases, a used test strip 
was removed and reinserted. 
Turn the meter off and on, or take 
the battery out for 20 seconds 
and reinsert it. Perform a blood 
glucose or control test.

The temperature is above or below 
the proper range for the system. 
Refer to the test strip package 
insert for system operating 
conditions. Move to an area with 
the appropriate conditions, wait 
5 minutes, and repeat the blood 
glucose or control test. Do not 
artifi cially heat or cool the meter.

The battery is almost out of power. 
Change the battery now. If the 
message reappears after the 
battery has been replaced, remove 
the battery again, press any meter 
button, then reinsert the battery.

The time and date settings may 
be incorrect. Make sure the time 
and date are correct and adjust, 
if necessary.

Blood glucose is below the defi ned 
hypoglycemic (low blood glucose) 
level. Repeat test.