248
Display Shows Possible Cause(s)
Possible Solution(s)
hI bG Warning
j
Your blood sugar may be
higher than the measuring
range of the system.
j
If you are experiencing any of the common
symptoms of high blood sugar, contact your
healthcare professional immediately.
j
Treat your high blood sugar as recommended by
your healthcare professional.
j
consider checking bG, ketones, and insulin.
j
See the pump user Guide for additional solutions.
Incorrect Pump
Time/Date
j
Pump time/date incorrect.
j
Set time/date on the pump (see pump User Guide).
Infrequent Pump
communication
j
At least two weeks have
passed since the meter and
pump have communicated.
j
ensure the meter and pump have
Bluetooth
wireless
technology turned on.
j
ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
j
It is important to use
Bluetooth
wireless technology
communication between the meter and pump
regularly if you utilize bolus advice.