Technical specification
Environmental information:
Condition
Temperature
Range
Relative
Humidity
Atmospheric
pressure
Operating
+10°C to +40°C
+50°F to +104°F
30% to 75%
(Non-
condensing)
700 hPa to
1060 hPa
Transport/
storage
-20°C to +50°C
-4°F to +122°F
If the CommunityBed Bed Lever is stored in conditions
outside the normal operating range, it should be
allowed time to stabilise, in normal operating
conditions, before use.
4.
Positioning
When selecting a position for the CommunityBed Bed
Lever, very careful consideration must be given to the
fitting zones shown below.
Zone A
– DO NOT fit the CommunityBed Bed Lever in
this zone. Fitting in this zone may cause an entrapment
risk resulting in serious injury
Zone B
– ONLY fit the CommunityBed Bed Lever in this
zone.
Zone C
– DO NOT fit the CommunityBed Bed Lever in
this zone. Fitting in this zone may cause an entrapment
risk resulting in serious injury
Community Bed Bed Lever Fitting Zone:
Description
Measurement
Zone A
(Measured from Headboard)
0 – 615mm
0 – 24in
Zone B
(Total length available)
984mm / 38.7in
Zone C
(Measured from Footboard)
0 - 395mm
0 – 15.5in
Description
Value
Overall dimensions
245mm W x 527mm H x 75mm D
Fits bed frame bar size
25mm W x 40mm H
Weight
2.3kg
Part number
STLEV-0-CB1-000
Fits bed
NSB-0-CB1-000
Service life
5 years
WARNING
The CommunityBed Bed Lever must be
positioned with due consideration to the best
practice described below. Failure to do so
may result in entrapment and serious injury.
Never use more than one CommunityBed Bed
Lever on each side of the bed.
Zone A
- Headboard
Zone C
- Footboard
Zone A
Zone C
Zone B