background image

 

 

13 

 

2.

 

Forward Tilt

 

 The chair can tilt forwards from 7 

degrees normal tilt to 10 degrees of forward tilt 
with lock bolts removal.  
 

3.

 

Legrest

 

 The legrest can be adjusted between 75 

degrees and 160 degrees; see page 5 for a 
diagram of this. The Legrest lever (Fig 27, Item 3) 
can be used to adjust this angle

 

Backrest Adjustment 

2.

 

Make sure the brakes on all 3 locking castors are 
locked on. 

3.

 

Locate the locking lever (Fig 28, Item 2) on the side 
of the backrest lockout slide (Fig 28, Item 1). 

4.

 

Release the backrest lockout slide by lifting the 
locking lever away from the backrest lockout slide.  
The backrest can now be adjusted using the 
backrest lever. 

5.

 

Pull the locking catch on the backrest lever, then 
pull the main lever (Fig 27, Item 1). Move the 
backrest to the desired measurement.  The 
backrest angle measurement can be read off the 
scale at the top of the backrest adjustment slide 
(Fig 28, Item 3). 
 

 

6.

 

Release the backrest adjustment lever to secure 
the backrest at the desired angle. 

7.

 

If locking out the backrest is required, return the 
locking lever to the locked position. If the locking 
lever is too loose or too tight it may be necessary 
to rotate the whole lever clockwise or anti-
clockwise respectively to achieve the desired 
locking action. 

8.

 

If locking out the backrest is not required there is a 
Velcro strap that can be used to hold the lever in 
the unlocked position 

 see Fig 29. 

 

 

 

Tilt-in-Space Angle Adjustment 

1.

 

Make sure the brakes on all 3 locking castors are 
locked on. 

2.

 

Pull and hold the tilt-in-space lever (Fig 27, Item 2) 
and move the chair to the desired tilt-in-space 
angle. 

3.

 

Release the lever to lock the tilt-in-space angle. 

 

Forward Tilt Adjustment 

This function enables some patients to exit the chair 
more easily, however it must only be used following a 
risk assessment of the patient.   

 

1.

 

The chair is delivered with this function locked. 
Remove the lock bolts on both LHS and RHS side 
on the chair using an Allen key. Ensure the correct 
bolts (Fig 30, Item 1) are removed.  

2.

 

Retain the lock bolts in case they need refitting in 
the future. 

Figure 28 

Figure 29 

WARNING 

Footplate must be removed before using the 
forward tilt function. 
 

Summary of Contents for CHAIR-0-SC3-050

Page 1: ...Configura Advance Chair Instruction Manual IFU SC3 001US Rev 02 March 2021...

Page 2: ...fied as a Class 1 Medical Device in accordance with the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation to this medical device affecting the user or the pati...

Page 3: ...pump cable could lead to dangerous electrical hazards if squeezed or crushed between moving parts The chair must not be used if there is any visible damage to any cables 12 Do not move the chair when...

Page 4: ...tended for use in primary healthcare environments where supervision can be provided if necessary for example long term residential care nursing homes or convalescent units The Configura Advance chair...

Page 5: ...1 050 Dropdown sidepieces and covers set of 2 DRPARM 0 SC1 250 Dropdown armrests 400 550mm 16 22 set of 2 DRPARM 0 SC1 050 Dropdown armrests 350 500mm 14 20 set of 2 DRPARM 0 SC1 150 Dropdown armrests...

Page 6: ...with the clevis at the back of the main frame Fig 3 Item 1 Ensure the Tilt in Space operating cable Fig 3 Item 4 is routed between the backrest and the backrest actuator assembly 6 Insert the clevis p...

Page 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Page 8: ...te Height range The footplate height can be adjusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers 5 Chair Seat Height range The Advance chair frame can be adjusted to give seat...

Page 9: ...h the other hand The depth setting can be read off the scale on the back part of the seat frame Fig 14 Item 1 Figure 13 5 Release the depth adjustment lever and move the backrest in and out slightly t...

Page 10: ...footplate See section 8 and place the chair carefully on its front 4 Using a 13mm spanner and 5mm Allen key undo and remove the Nylock nut and shoulder bolt Fig 19 Item 1 and Item 2 Keep the bolt was...

Page 11: ...ig 24 Items 1 2 into the two sockets either side of the footplate height adjustment frame Fig 21 Items 3 4 Push the footplate into the tubes until the plungers latch with an audible click 3 Ensure the...

Page 12: ...can be used to restrict the use of the backrest angle adjustment 1 Tilt in Space The tilt in space position can adjust the seat angle from 7 degrees to 37 degrees from the horizontal position see pag...

Page 13: ...e 7 If locking out the backrest is required return the locking lever to the locked position If the locking lever is too loose or too tight it may be necessary to rotate the whole lever clockwise or an...

Page 14: ...to press Care must be taken to ensure the castor brakes are always locked when the chair is in use being assembled or dismantled so that the chair does not move accidentally Individually braked 125mm...

Page 15: ...r can be braked at the same time using the foot lever 1 To lock the central brake press the rear red part of the foot lever down until the brake locks Figure 31 shows the brake locked 2 To unlock the...

Page 16: ...s profiled headrest external laterals 3 Clean as described in the Cleaning Information 13 Troubleshooting Problem Possible solution s Backrest will not recline manual version Ensure the backrest locko...

Page 17: ...tplate securely attached 12 Are the plungers securing the sidepieces locked and secure 13 Is there any risk of entrapment or patient injury 14 Are any accessory electrical cables pinched crushed or da...

Page 18: ...t IPX4 Degree of protection against liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in acco...

Page 19: ...19 This page intentionally left blank...

Page 20: ...20 IFU SC3 001US REV 02...