OMNISTIM
®
FX
2
CYCLE / WALK PROFESSIONAL USER MANUAL
11
COPYRIGHT © 2011 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Factory Settings
The
OMNISTIM
®
FX
2
CYCLE / WALK
comes with the factory settings defaulted to the CYCLE program. This is
generally suitable for most patients using the Omnicycle
®
or Omnicycle
®
Elite. The following chart shows the
available channel configurations.
PROGRAM
PROTOCOL
USER ADJUSTABLE PARAMETERS
RPM
PHASE DURATION
TREATMENT TIME
CYCLE
Variable
Range = 1 to 75 rpm
Range = 50 -120 µs
Range = 1 to 30 min
WALK
Variable
DEFAULT
PROGRAM
DEFAULT
RPM
PHASE DURATION
TREATMENT TIME
PROGRAM BUTTON A
CYCLE
45 rpm
100 µs
15 min
PROGRAM BUTTON B
WALK
60 rpm
Professional Programming Sequence
1.
Press the on/off button (1) to power on the device.
2.
At power on, the display will default to the factory loaded CYCLE
program This is a two channel program requiring output adjustment of
both channel A (4) and channel B (5). It may be necessary to change the
default parameters and/or settings to meet the patient’s specific needs.
3.
To change programmable settings for the CYCLE program, press and
hold the Program Button A (2) for 5 seconds to enter the programming
mode. In the programming mode, the program setting for RPM will
flash. Use the (+) (-) buttons (4) to increment or decrement the adjustable
setting. When set, press the Program Button A (2) once to proceed to the
next programmable setting.
4.
To set the device for the WALK program, press the Program Button B (3) once to select WALK then press and
hold for 5 seconds to enter the programming mode. In the programming mode, the program setting for RPM will
flash. Use the (+) (-) buttons (5) to increment or decrement the adjustable setting. When set, press the Program
Button B (3) once to proceed to the next programmable setting.
5.
Once complete, press the on/off button (1) to power down the device. The new settings will be saved as the new
“default program” that starts the next time the device is powered on.
Note:
The settings will not be saved unless
the unit is powered down following programming.
NOTE
:
The above settings cannot be accessed during operation and are only available following the start up
screen. All changed settings will be saved for the next use. To reset to factory settings, press and hold both
Program Buttons A + B for 5 seconds.
Program Counters
1.
The
OMNISTIM
®
FX
2
CYCLE / WALK
is equipment with imbedded counters to tell you how many times the
patient has used the device.
2.
Press the on/off button (1) and hold for 5 seconds to enter the counter readings.
3.
The unit will default to the CYCLE program first, and will display a 3 digit number to represent the number of
times that the program has been used.
4.
Use the (+) (-) buttons (4) or (5) to toggle to the WALK counter. It will be displayed in the same 3 digit number
as described above.
5.
To clear the counters and reset to zero (0), press and hold both Program Buttons A + B for 5 seconds.
NOTE
:
When you reset the counters, this will also rest the unit to factory settings. You will need to reprogram
any patient specific programming necessary for the patients treatment plan..