12 NEUROPROBE
®
500 PRO USER MANUAL
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
7.
LASER ON or OFF activates the additional LASER parameter selections and dosage options. This mode should
be used only with the Neuroprobe
®
500 Pro optional LASER probes and is covered in the LASER probe
operator manual supplement.
NOTE:
The factory settings cannot be accessed during operation and are only available following the start up
screen.
Power Supply Operation
Neuroprobe
®
500 Pro is a line power only device. It is not intended for use with batteries.
When the Power Supply is plugged into the unit, the LED on top right of the Neuroprobe
®
500 Pro illuminates and
stays on as long as the Power Supply is plugged into the unit and connected to the AC power.
OPERATIONAL SEQUENCE
Infrared Therapy with SLD Pads
Select the dual rectangular, knee, elbow, neck, back or lower extremity SLD Infrared Pad. Plug it into the
probe output socket.
Apply the SLD Infrared Pad to the treatment area. Use an infection control cover over the SLD Infrared
Pad and proper infection control procedures. Use the Velcro
®
straps supplied on the pad or use separate
straps to gently place the SLD Infrared Pad (s) over the treatment area. Position the pad (s) so that the RED
SLD are placed over the key nerves, arteries and acu/trigger points for treatment of the knee, back or lower
extremity based on the pad used.
Press the POWER button (1/5) to turn on the Neuroprobe
®
500 Pro. The display screen will illuminate, and
the unit will be in the SELECT
MODALITY/PROGRAM mode.
Select the Infrared Therapy treatment program by using the MODALITY button (2). Note that although the
SLD PADS will function in probe modes, the detection and electro-stimulation feature of the probe will not
be available as the SLD INFRARED PADS plug into the same output socket as the probe.
To start the treatment, push the START/STOP button (3) once. The following screen will be displayed.
INFRARED THERAPY T15