26
NEUROPROBE
®
500 PRO USER MANUAL
COPYRIGHT © 2006 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Waveform:
Burst Alternating Current
Output:
0 mA
0
– 1.8 mA
II. SENSORY
– MOTOR, IFC AND SLD INFRARED PAD PROTOCOLS
Electro therapy may be applied with medium frequency currents to provide greater tissue penetration and higher
comfort. Electro therapy with medium frequency current is applied using bipolar electrode placements whereas IFC
current is applied by crossing the electrodes over the desired treatment area. The current is passed through the
treatment site by creating a third vector field from the crossing or “interfering” of the two channels in the tissue.
The purpose of interferential current therapy is to provide deep tissue treatment, which is not generally obtainable
with conventional electrotherapy approaches. Its primary application is in the reduction of pain and in the
stimulation of increased blood flow in the deeper tissues and muscles.
The following pads programs are available in Sensory - Motor mode:
PADS SENSORY:
PARAMETER
DEFAULT VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
5.0 KHz
2.0 KHz, 2.5 KHz,
4.0 KHz, 5.0 KHz or
10 KHz
Burst frequency:
100Hz
.0 Hz to 999 Hz
Burst rate scan:
20%
0
– 50%
Burst scan time:
10 seconds
0
– 20 Seconds
Beat:
Sweep:
Rate Scan Time:
Modulation:
ON
ON or OFF
Vector:
OFF
OFF, Fast 90º, Slow 90º,
Fast 45º, Slow 45º
Auto Intensity:
Treatment time:
15 minutes
0
– 99 Minutes
Waveform:
Burst Alternating Current
Output:
0 mA
0
– 99 mA
PADS MOTOR:
PARAMETER
DEFAULT VALUES
RANGE OF USER
ADJUSTABLE VALUES
Carrier frequency:
2.5 KHz
Burst frequency:
8.5 Hz
.0 Hz to 999 Hz
Burst rate scan:
50%
0
– 50%
Burst scan time:
10 seconds
0
– 20 Seconds
Beat:
Sweep:
Rate Scan Time:
Modulation:
ON
ON or OFF
Vector:
OFF
OFF, Fast 90º, Slow 90º,
Fast 45º, Slow 45º
Auto Intensity:
0%
0
– 20%