Calibration Key
Tab. 3–4: Calibration Grid Keys
If the Calibrator lot is different than the current calibrator material being used for calibra-
tion, proceed with the following as described in
4.2. Target values
From the Calibration grid, select the «Target» key (see
Tab. 3–4: Calibration Grid Keys,
3-26) to open the «Calibration Target» screen for editing.
Fig. 3–24 Editing the target values
Select the «Edit» key to edit the target fields.
Read the barcode label on the calibrator with the external Barcode Reader or type in all the
calibrator information using the calibrator Assay sheet that comes with the calibrator kit.
Once all entries have been made, select the «OK» key to save the information
Key
Name
Function
Automatic Calculation
Runs an automatic calculation of the new
coefficients (see
page 3-28). This key is enabled
only if a minimum of 5 runs on calibrator
has been performed
Target
Access to the Target values window
(see
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...