Maintenance
& Troubleshooting
Super User menu description
Pentra
80 - User Manual - RAB108EA
7–35
5.3. Others
Fig. 7–42 Super User\Others
1-
Cycles Counter
: shows the number of specified cycles performed by the
instrument.
2-
Run Park Syringe Position
: This key allows the user to place the syringe pistons in
a safe position for long-term non-usage or storage and when the instrument will be in
transport from one location to another.
3-
Run Maintenance Carriage Pos.
: This key allows the user to automatically move
the Sample Carriage over the chamber area for sample probe replacement and/or other
maintenance procedures that may require the movement of the sample carriage.
4- Each
Calibration Coefficient
can be forced to a value:
– Enter the value you want to modify in the «Calibration coefficients» grid.
– Confirm the modifications by pressing the «Accept values». A notification is
automatically done in the «calibration» logs
Any modification of the coefficients will affect the results.
4
2
3
1
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...