Pentra
80
7–30
Pentra
80 - User Manual - RAB108EA
5.2. Hydraulical menu
5.2.1. Drain chambers
Fig. 7–35 Super User\Hydraulical\Drain Chambers
1-
Rinse
: Drains the rinse chamber
2-
HGB
: Drains the HGB chamber
3-
LMNE
: Drains the LMNE chamber
4-
RBC/PLT
: Drains the RBC/PLT chamber
5-
WBC/BASO
: Drains the WBC/BASO chamber
6-
All
: Drains all the chambers.
5.2.2. Prime cycles
This menu will allow the user to prime reagents into the instrument.
Run this procedure after service has been performed on the instrument.
This function allows only the priming of reagents into the instrument. It does not
set the reagent quantities to 100%
4
5
6
1
2
3
Summary of Contents for Pentra 80 RAB108EA
Page 1: ...User Manual P n RAB108EA ABX Diagnostics B P 7290 34184 MONTPELLIER Cedex 4 FRANCE...
Page 12: ...10 Section9 Glossary 1 Glossary 9 2 2 Index 9 4 3 9 11...
Page 87: ...Quality Assurance Calibration Pentra 80 User Manual RAB108EA 3 31 Fig 3 29 Delete options...
Page 162: ...Pentra 80 5 10 Pentra 80 User Manual RAB108EA Fig 5 6 Units tab...
Page 194: ...Pentra 80 5 42 Pentra 80 User Manual RAB108EA...
Page 214: ...Pentra 80 6 20 Pentra 80 User Manual RAB108EA...
Page 264: ...Pentra 80 7 50 Pentra 80 User Manual RAB108EA...
Page 284: ......
Page 286: ......
Page 288: ......
Page 290: ......
Page 292: ......
Page 294: ......
Page 296: ......
Page 306: ...Pentra 80 9 10 Pentra 80 User Manual RAB108EA Z Zoom Chap 4 28...