13
SECTION 2 INSTALLATION
2E. Remote Control
4-20mA On/Off Power Remote Control
• Remote control of ozone production level can be
achieved by 4-20mA at 10 V signal. The function of this
power control is not linear.
• Remote On/Off. Remote Switch must be normally open
for the generator to be “On” and closed to stop the
Ozone production.
• That switch has to be connected to dry contacts only
(not connected to ground or connected to any voltage
source AC or DC)
• Remote Ozone Generator status indicator.
a) During Normal Ozone generator operation. Pins 5 and
6 are normally open, pins 6 and 7 are normally closed.
b) If Remote Control Switch is Off or safety shutoff is ac-
tivated the circuit between pins 5 & 6 will be closed and 6
& 7 open.
Summary of Contents for NANO 15