PROBLEM: 

POSSIBLE CAUSE / SOULTIONS: 

Moisture collects inside 

 

Door gasket is not sealing properly. Check for debris, 
cracks, and items passing through door at the gasket.                                                             

 

The refrigerator or freezer is facing a doorway or is 
underneath of air conditioning vent. Relocate the unit or 
redirect air vent. 

 

Too many door openings. Minimize time door is open. 

 

Hot, humid weather increases condensation. 

 

Make sure there is a water trap (U-shaped loop) in the 
drain tube near the compressor. This will “trap” a small 
amount of water in the loop and prevent air from 
entering the chamber through the tube.  

Moisture collects on outside 
surface 

 

Hot, humid weather increases condensation. 

 

As humidity decreases, moisture will disappear. 

The display shows “SO” and 
the compressor does not run 

 

The temperature sensor is open. Call for service.                                                                                                                                                                                                                                                                                                                                                       

The display shows “SC” and 
the compressor does not run 

 

The temperature sensor is shorted. Call for service.  

The display shows “HH” and 
the compressor does not run 

 

The ambient temperature is above 40C/104F. Stop using 
immediately. 

The display shows “LL” and the 
compressor does not run 

 

The ambient temperature is below -40C/-40F. Stop using 
immediately.  

MOISTURE DURING THE SUMMER SEASON 

The amount of moisture, condensation, or high humidity related issues increase during the summer 
and,  in  most  cases,  will  self-resolve  when  the  weather  cools  down.    Please  note  a  refrigeration 
system does NOT generate moisture or water but simply condenses the moisture that is already in 
the chamber. Keeping the unit in an air conditioned, low humidity space will resolve many issues. 
Other things you should check 

1.

 

Location of the refrigerator (See Quick Troubleshooting Guide above) 

2.

 

Door sealing and frequency of door opening event (See Quick Troubleshooting Guide above) 

3.

 

Make  sure  there  is  a  water  trap  (U-shaped  loop)  in  the  drain  tube  near  the  end.  This  will 
“trap”  a  small  amount  of  water  in  the  loop  and  prevent  air  from  entering  the  chamber 
through the tube.  

 
BEFORE CALLING THE MANUFACTURER’S TECHNICAL SUPPORT DEPARTMENT, 

please have the 

unit’s model and serial number ready as well as the problem description.  The model and serial 
number  is  located  on the  serial tag  which  can  be  found  on  the  interior  left upper  wall  of the 
unit. 

Summary of Contents for ABT-RFC-7P

Page 1: ...e sure you get a signed copy Send a written request for an inspection to the carrier B Concealed Damage This type of damage may not be discovered until the unit is being unpacked When concealed damage...

Page 2: ...L SERVICE please have the unit s model and serial number ready as well as the problem description The model and serial number is located on the serial tag which can be found on the interior left OR ri...

Page 3: ...y receipt acknowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along wi...

Page 4: ...is cabinet must conform to all National and Local Electrical Codes Consult the cabinet Serial Data plate for voltage cycle phase and amp requirements before making connection VOLTAGE SHOULD NOT VARY M...

Page 5: ...ecially if the weather is very hot This is normal The motor compressor will cycle normally as soon as the excess heat has been removed TEMPERATURE CONTROL NOTE The control is preset at the factory to...

Page 6: ...ides of the unit to allow for proper air flow and therefore more even cooling of the product When loading the top shelf avoid blocking the evaporator fan s if installed There should be at least 4 inch...

Page 7: ...ides the following functions Red LED for visual indication of an alarm condition Beeper for audible indication of an alarm condition A preprogrammed 5 minute delay from the time a set point is attaine...

Page 8: ...Pon The control keys are now unlocked and you can perform the programming functions below Adjusting the temperature control set points NOTE The temperature control set points are preset at the factor...

Page 9: ...isplay Press the UP or DOWN keys several times to cycle through the parameters until CF is displayed Press SET to display the current setting either C for Celsius or F for Fahrenheit Use the UP or DOW...

Page 10: ...s Technical Support Department before making any adjustments to determine if adjustment is necessary and if so to make sure it is performed properly This unit has a manual thermostat which controls th...

Page 11: ...the temperature transient while loading Other tips for successful loading include Leave about 2 3 inches of free space along the back and sides of the unit to allow for proper air flow and therefore...

Page 12: ...d the proper refrigerator air circulation For the proper air circulation place the products evenly on each shelf Do not push against the refrigerator s rear or side walls Evaporator is blocked by fros...

Page 13: ...bove 40C 104F Stop using immediately The display shows LL and the compressor does not run The ambient temperature is below 40C 40F Stop using immediately MOISTURE DURING THE SUMMER SEASON The amount o...

Page 14: ...e the proper tools When cleaning stainless steel surfaces take care to use non abrasive tools Soft cloths and plastic scouring pads will not harm the steel s passive layer Stainless steel pads may be...

Page 15: ...er Sheen Passive Film Breakdown Corrosion on metals is everywhere We recognize it quickly on iron and steel as unsightly yellow orange rust Such metals are called active because they actively corrode...

Page 16: ...e paid by the customer ADDITIONAL THREE YEAR COMPRESSOR WARRANTY In addition to the two 2 year warranty stated above Horizon Scientific Inc warrants its hermetically and semi hermetically sealed compr...

Page 17: ...t damage during transit or installation fire flood or other external causes It does not apply to defects resulting from failure to properly install operate or maintain the product in accordance with t...

Reviews: