If excessive resistance is felt while advancing the Thumb Advancer, pull back the Thumb Advancer.
•
Slide the Safety Release forward to collapse the Locator Wings.
•
Check that the number
“2”
on the Plunger exits the number window completely.
•
Gently push down on the patient’s body with the left hand. Remove the Clip Applier with the right hand without rocking or twisting.
Use of Safety Release
StarClose SE
Vascular Closure System
12