StarClose SE

Vascular Closure System

Deployment Steps

©2011 Abbott. All rights reserved. ANZ 080091 10/11

Summary of Contents for StarClose SE

Page 1: ...StarClose SE Vascular Closure System Deployment Steps 2011 Abbott All rights reserved ANZ 080091 10 11...

Page 2: ...Button Side arm Slot Finger Loop Stabilizer Access Port Access Port Safety Release Plunger Thumb Advancer Finger Post Finger Post Finger Post Clip Delivery Tube StarClose SE Vascular Closure System 1...

Page 3: ...StarClose SE Vascular Closure System Clip Delivery Tube A Clip B Clip Delivery Tube C Sheath Splitter A B C 2 2011 Abbott All rights reserved ANZ 080091 10 11...

Page 4: ...rClose SE Vascular Closure System Clip Material Nitinol Nickel Titanium Alloy Actual diameter of the closed Clip 4 mm Clip thickness 0 2 mm 4 mm 0 2 mm 3 2011 Abbott All rights reserved ANZ 080091 10...

Page 5: ...StarClose SE Vascular Closure System Exchange System J Tip Guide Wire Sheath Tube Exchange Sheath Hub Dilator 4 2011 Abbott All rights reserved ANZ 080091 10 11...

Page 6: ...ery Profunda artery External iliac artery Inferior epigastric artery Inguinal ligament Superficial femoral artery Perform femoral angiogram Re Prep groin Perform nick and spread External iliac artery...

Page 7: ...e Flex guide excessively during insertion Hold the bottom of the Hub with the left hand Firmly push the Hub into the Clip Applier A CLICK should be heard Gently pull on the Hub to confirm engagement C...

Page 8: ...meters Using the right thumb depress the Plunger firmly to deploy the Locator Wings which advances the Thumb Advancer about 3 centimeters The initial splitting of the Sheath will be visible above the...

Page 9: ...Wings are against the arterial wall and slight resistance is felt While maintaining the Locator Wing position against the arterial wall push the Thumb Advancer forward with the pad of the right thumb...

Page 10: ...ntly push the Clip Applier down to seat the Clip Delivery Tube on top of the access site Pulsation of the artery may be felt While maintaining downward pressure depress the Deployment Button with the...

Page 11: ...depress the area around the puncture site to express any blood in the tissue tract Test for security of the close by having the patient lift their head off the table and cough and lift their leg again...

Page 12: ...Arterial Perspective StarClose SE Vascular Closure System CLICK 1 CLICK 2 CLICK 3 CLICK 4 11 2011 Abbott All rights reserved ANZ 080091 10 11...

Page 13: ...apse the Locator Wings Check that the number 2 on the Plunger exits the number window completely Gently push down on the patient s body with the left hand Remove the Clip Applier with the right hand w...

Page 14: ...Thumb Advancer exits the number window completely Pull back the Thumb Advancer Slide the Safety Release to collapse the Locator Wings Check that the number 2 on the Plunger exits the number window com...

Page 15: ...hysician Prior to use it is important to read the package insert thoroughly for instructions for use warnings and potential complications associated with use of this device Information contained herei...