SOTOX A™ MOBILE ANALY ZER | USER GUIDE
6 | EN
1. INTRODUC TION
• Please ensure that the mains supply cable for the power supply unit is appropriately rated and approved as per the applicable
national requirements.
• As with any electrical instrument or equipment, care is required when handling the analyzer to ensure that it is not subjected to heavy shocks or
dropped from any height. Never exert undue force on the surface of the display, or expose any part of the analyzer to sharp or abrasive objects.
• Use the analyzer indoors when connected to A/C power.
• Do not open the analyzer. The analyzer contains no user serviceable parts. Any unauthorized opening of the analyzer can cause hazards and
will invalidate warranty claims.
• It is highly recommended that all PCs supporting the Software Application Suite (if used) carry up-to-date antivirus software.
• The test cartridges are designed for single use only – do not attempt to re-insert a used test cartridge into the analyzer.
• Do not insert fingers or objects other than the test cartridge into the test cartridge slot. This may result in harm to the user or damage to
the analyzer.
• Do not use dirty or damaged test/QC cartridges as this may damage the analyzer or cause an invalid test result.
• Do not use the analyzer or any of its electrical accessories if these appear to be damaged. Contact Customer Service to arrange for
replacement parts.
• Do not use the analyzer near a magnetic source.
• Do not ingest the silica gel packs.
• Do not open power supply casing or printer.
• Allow the analyzer to reach ambient temperature before use. If moving the analyzer between storage and operating environments of
different temperatures or humidities, this may take at least 30 minutes. If climatic differences are significant, allow more than 30 minutes for
the analyzer to acclimatise. It is the responsibility of the user to ensure the date and time settings are correct.
1.9
Analyzer calibration
The SoToxa™ Mobile Analyzer is factory calibrated and does not require any further calibration by the user.
However, it is recommended to carry out a regular performance check using the quality control test cartridges provided with the SoToxa™
Mobile Test System (see section 7 QC Tests).
1.10 Maintenance and cleaning
If required, the exterior of the SoToxa™ Mobile Analyzer and its accessories can be wiped with a lint free cloth or a damp cloth with mild
detergent. Wipe away any spillages immediately. Do not use abrasive cloths or other material, and do not allow the unit to come into contact with
any potentially corrosive substances. If using a damp cloth, ensure the analyzer is turned off and disconnected from mains A/C power.
DO NOT try to clean inside the analyzer or open the analyzer for cleaning.
The analyzer does not contain any user serviceable components. If a fault is suspected or routine factory maintenance is required, then please
contact Customer Service.
1.11 Servicing
The SoToxa™ Mobile Analyzer requires an annual service which must be carried out by a qualified service engineer. It is important to
have your analyzer serviced annually to ensure it is performing correctly. This requires the analyzer and accessories to be returned to the
manufacturer or authorised servicing facility for a short period of time.
The date that the next service is next due can be viewed in the analyzer summary screen, and the analyzer will automatically display the
service symbol in the status line when service is due in less than one month. If the analyzer is not returned to the manufacturer for servicing
within this time, the symbol will turn from white to orange.
Service status (OK, due or overdue) will be displayed in the test details for each test and will also be printed on test printouts.
Accuracy of tests carried out on an analyzer which is overdue a service, may be reduced. Failure to return the analyzer for annual service will
invalidate warranty claims.