
12
5.
Use the clinician programmer app to communicate with the IPG and perform intraoperative
testing to confirm that the system is operational. See the clinician's manual of the clinician
programmer app for instructions.
NOTE: IPG output may not be identical to that of the trial stimulator at the same settings.
6.
Ensure that the IPG is away from the pocket incision suture line, close the pocket incision,
and apply the appropriate dressings.
Replacing the IPG
The following steps outline the suggested procedure to replace an IPG:
1.
Turn off stimulation or verify that it is turned off.
CAUTION: Exercise care when using sharp instruments or electrocautery around leads
or extensions, or they may be damaged.
2.
Open the IPG implant site per normal surgical procedure.
3.
Insert the torque wrench through the septum of the IPG header and loosen the setscrew by
turning it counterclockwise.
CAUTION: When performing the following step, do not bend the lead or extension
sharply; or it may be damaged.
4.
Gently remove the lead or extension from the IPG header; then clean and dry all
connections, ensuring they are free of fluid and tissue.
5.
To complete the IPG replacement procedure, see the following sections: “Connecting a Lead
or Extension to the IPG” (page 9) and “Implanting the IPG” (page 11).
Disposing of Explanted Components
Explanted Abbott Medical components should be returned to Abbott Medical for proper disposal.
To return an explanted component, place it in a container or bag marked with a biohazard label
and coordinate the return with your Abbott Medical representative or Technical Support.
Checking the Status of the IPG Battery
The IPG contains a nonrechargeable battery. The amount of time that the battery will provide
active stimulation depends on the patient’s stimulation settings and daily usage time. To check
the status of the IPG battery, use the clinician programmer app or patient controller app. For more
information about this function, refer to the clinician’s programming manual and the user’s guide
for the patient controller app. For information about estimating longevity of the IPG battery, see
the appropriate appendix in this manual.
NOTE: IPG battery status is available one day after first using the clinician programmer app
to program the IPG.
The following list provides general information about the battery status:
A low-battery warning will appear on the clinician programmer app or patient controller app
when the battery is approaching its end of service.
Stimulation will automatically stop when the battery cannot support stimulation.