Table 5. Guidance and manufacturer's declaration - electromagnetic immunity (conducted RF
and radiated RF)
Test
IEC 60601
Test Level
Immunity
Compliance Level
Electromagnetic Environment -
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.7 GHz
(3 Vrms)
[V1=3]
3 V/m
[E1=3]
Portable and mobile RF
communications equipment should be
used no closer to any part of the
Merlin™ 2 PCS, RF Antenna, and
Merlin PSA, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
80 MHz to 800 MHz
800 MHz to 2.7 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey*, should
be less than the compliance level in
each frequency range.** Interference
may occur in the vicinity of equipment
marked with the following symbol:
*Electromagnetic site survey- Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
9
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
10
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.