3-5
Questions? Refer to Page 11-22 For Technical Support Instructions.
How to Obtain an Arterial Blood Sample (For Glucose Testing):
• Use a new pair of clean gloves for each patient.
• Use only fresh whole blood samples.
• Clear the arterial line before drawing a blood sample into a heparinised syringe.
• Use the sample within 30 minutes of collection.
• Mix the blood specimen well immediately before applying the sample to the target area of the test
strip.
• Allow a drop of blood to form at the tip of the syringe.
• Apply a drop of blood to the target area at the end of the test strip.
How to Obtain a Neonatal Capillary Blood Heel Stick Sample (For Glucose Testing):
• Use a new pair of clean gloves for each patient.
• Use only fresh whole blood samples.
• Collect the capillary blood using an auto-disabling, single use lancing device and an appropriate
technique.
• Apply a drop of blood to the target area at the end of the test strip.
Chapter 3: Patient Test
Summary of Contents for FreeStyle Precision Pro
Page 1: ...Operator s Manual for Healthcare Professionals...
Page 88: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test Ketone Beta Hydroxybutyrate...
Page 98: ...7 Linearity Test...
Page 135: ...9 Review Setup for Wireless Communication Optional Meter Wireless Configuration...