Table 10. Technical specifications
Output voltage
5 VDC
Output current
up to 1 A
Power adapter
certification marking
Power adapter should have one or more of the following
certification markings:
Appendix D: Electromagnetic Compatibility
Guidelines
The charging system (Model 16000), hereafter the device, is medical equipment and
should be used with the following guidance.
The device requires special precautions with regard to electromagnetic compatibility
(EMC) and should be used in accordance with the information provided in this manual.
The device has essential performance, as defined by
IEC 60601
‑
1:2005+AMD1:2012+AMD2:2020 (Ed 3.2). Performance of the device was
maintained during electromagnetic disturbance testing.
The device is intended for use in the electromagnetic environment specified in the
following tables. The user should ensure that it is used in such an environment.
CAUTION:
▪
The device complies with the limits for medical devices contained in
▪
IEC 60601
‑
1
‑
2:2014+AMD1:2020 (Ed 4.1) and
CISPR11:2015+AMD1:2016+AMD2:2019 (BS EN 55011:2016+A2:2021). However,
the device may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to mitigate this effect by reorienting or
relocating the receiving device.
▪
To avoid increasing emissions or decreasing immunity from a device or system,
▪
use only components approved by Abbott Medical with this system. Do not use
Abbott Medical components with devices or systems that are not approved by
Abbott Medical.
Table 11. Guidance and Manufacturer's Declaration - Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environment
Guidance
RF emissions
CISPR 11
Group 1
The device produces RF emissions as
a result of its internal and system
interface functions. Its RF emissions
are very low and are not likely to
cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-voltage
Harmonic emissions IEC 61000-3-2
Class A*
23
▪
▪