Installation Procedures and Special Requirements
Printer Installation
Section 2
To inspect the Sample Loader:
1. Unpack the instrument from its shipping container and place it on a flat
surface.
2. Remove the protective plastic.
3. Inspect the Sample Loader for signs of obvious damage and verify the loader
is securely attached to the Analyzer.
4. Remove the set of 5 tube racks from the Sample Loader Accessory Kit.
5. Inspect all the racks for signs of damage.
6. Verify that each rack has a bar code label in position 1 and that the label is on
the same side as the open slots. Refer to Figure 2.9 for proper label placement
and alignment.
NOTE:
For the Sample Loader to operate properly, the racks must be
placed to the right of the tower (the “load” area) with the labels and
slots facing the operator, not the Analyzer, so that the Bar Code
Reader can read the rack label and the tube labels through the rack
slots. Refer to
information on the Sample Loader.
Figure 2.9 Tube Rack Showing Label Placement Locations
NOTE:
Tube bar code labels must be
≥
3 digits.
1 Rack Bar Code Label
2 Rack Tube Position
Labels
1
2