Section 10
Troubleshooting and Diagnostics
The Sample Loader beeps but does not begin processing
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The internal cable that connects the
Sample Loader to the Analyzer is
loose or disconnected.
1. Contact your Abbott Hematology
Customer Support Center.
2. Circuitry malfunction.
2. Press
[FAULT REPORT]
in the main
DIAGNOSTICS MENU
and record any
fault listed.
3. Turn the instrument OFF and ON
again and prime the instrument.
4. If the problem persists, contact your
Abbott Hematology Customer
Support Center.
The Display Monitor membrane touch keys are not operational
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The computer is performing a
function that inhibits the keys.
1. No action required. Wait for the
computer to resume the operation.
Refer to the screen for the current
Status Box message.
2. There is an incomplete operator
entry.
2. Complete the operator entry or press
the ESC key on PC keyboard.
3. A data transmission to the printer or
laboratory computer is in process.
3. No action required. Wait until the
transmission is completed.
4. Membrane keyboard or circuitry
malfunction.
4. Power OFF the Analyzer. Then
reseat the Membrane Keypad cable
on the rear panel of the Data Module
and power Analyzer ON.
NOTE:
For the location of the
Membrane Keypad
Interface connector, refer to
.
Table 10.4 Non-Functional Fault Conditions (Continued)