Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Table 10.4 Non-Functional Fault Conditions
Information for each message is listed in order from the most likely cause of the
message to the least likely cause of the message. Therefore, always troubleshoot
a problem in this order. If a problem cannot be resolved by the corrective action
indicated in this table, call for technical assistance.
Analyzer will not power ON
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. Power cord is not securely
connected to the Analyzer or is not
connected to the power outlet.
1. Ensure that the power cord is
securely connected to the Analyzer
and verify that it is connected to the
power outlet.
2. Power source is defective.
2. Turn the power switch OFF and
connect the system to a different
power source.Turn the power switch
ON.
3. Analyzer fuse has failed or is
incorrect.
3. Analyzer fuse is located above the
power cord connector on the rear
panel. Check the fuse as directed in
the
4. Defective power switch or other
system malfunction.
4. Contact your Abbott Hematology
Customer Support Center.