CELL-DYN
®
3200 System Operator’s Manual
9-55
9140181K—July 2002
Section 9
Service and Maintenance
Syringe Replacement
A syringe should be replaced under the following conditions:
1. There is evidence of leaking
2. It is suspected of being a source of imprecision.
Materials Required
1. Replacement syringe
2. Gloves, lab coat, and safety glasses
Procedure
1. The procedures for removing and installing the Sample Injection Syringe,
WBC Lyse Syringe, HGB Lyse Syringe, and Diluent/Sheath Syringe are
described earlier in this subsection.
2. After the old syringe has been removed, follow the instructions in
Syringe
Cleaning
to reinstall the new syringe and bring the Analyzer back to the
READY state.
Fuse Replacement
There are two conditions under which a fuse should be replaced:
1. If the fuse has failed, or
2. If the power setting is changed from 110/120 VAC to 220/240 VAC or vice
versa.
Materials Required
1. Replacement fuse (8-amp T (Slo-Blo) for instruments operating at
110/120 VAC or 4-amp T (Slo-Blo) for instruments operating at
220/240 VAC)
2. Flathead screwdriver
Procedure
CAUTION: Electrical Hazard.
Turn off the power to the instrument and
disconnect the power cord before removing any instrument panel that is
securely fastened in place by screws or prior to replacing fuses. Replace
only the externally accessible fuse located immediately above the power
cord connector on the rear of the instrument. Use replacement fuses only of
the specified type and electrical rating.
1. Turn the system OFF and disconnect the power cord. Locate the plastic fuse
holder on the Rear Panel. (Refer to Figure 9.22.)
2. With the system OFF, insert a flathead screwdriver into the notch on the fuse
holder.
3. Push in and turn the fuse holder counter clockwise to release the holder. Pull
the holder with the fuse completely out of its slot.
4. Pull the fuse to remove it from the holder. Check the fuse.