10-74
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Loader Fault 147: unload area hardware malfunction
/BL
The sensor system is indicating that the unload area is full (contains 5 racks), but
is not indicating that it is also nearly full (contains 4 racks).
– – – - Probable Cause – – –
– – – - Corrective Action –
1. One or more racks did not move
properly in the unload area.
1. Check for an obstruction that is
preventing rack movement in the
unload area and, if one is found,
remove it. Then, reset the racks and
reset the Sample Loader by
pressing the following keys in
order:
[CLEAR FAULT]
,
[START LOADER]
,
[RESET LOADER]
.
2. A failure of the sensor system or
related electronics has occurred.
2. Reset the racks and reset the
Sample Loader. If the message
appears repeatedly, contact your
Abbott Hematology Customer
Support Center.
Table 10.2 General Fault Conditions (Continued)