![Abbott CELL-DYN 3200 System Operator'S Manual Download Page 454](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985454.webp)
10-20
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Diagnostics Menu
Section 10
Serial Test
The
[SERIAL TEST]
key is used to test the functionality of the COM1 port (referred
to as the “serial interface connector”) at the rear of the instrument. This test is
designed to assist in troubleshooting problems with interfacing to a Laboratory
Information System (LIS). The loop-back connector must be connected to the
COM1 port before performing the test.
NOTE:
If the laboratory does not have a LIS, the loop-back connector may remain
connected to the COM1 port for convenience, as it does not interfere with
routine operation. If a LIS is usually connected, the loop-back connector
should be stored near the instrument when the connector is not in use.
Figure 10.19 Serial Test Screen
When the
[SERIAL TEST]
key is pressed, the following soft key labels (see
Figure 10.19) are displayed:
STOP TRANSMISS
TRANSMIT MESSAGE
RETURN
The screen displays the following:
Serial Interface Test
1.
See Interface Specification.
2.
If transmission in progress, press [STOP TRANSMISS]
key first.
3.
Attach Loop-back Connector to the serial interface
connector on back of the Data Station.
4.
Press the [TRANSMIT MESSAGE] key to start the test.