9-10
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Preventive Maintenance Schedule
Section 9
Preventive Maintenance Schedule
The following procedures should be performed at the scheduled time intervals or
more often, depending on usage, as determined by each individual laboratory.
Keep a record of routine maintenance and parts replacement to ensure proper
equipment care and to assist Abbott Customer Support Specialists and Abbott Field
Service Representatives in diagnosing instrument problems.
Daily
Run Auto-Clean Cycle
Clean Closed Sample Aspiration Needle
Clean Closed Sample Tower
Weekly
Clean Sample Loader and Racks
Check Sample Transfer Pump Tubing
Run the Extended Auto-Clean Cycle if routinely running reticulocyte counts.
Monthly
Clean Fan Filter
Run Extended Auto-Clean
Replace Diluent/Sheath Filter
Semi-Annual
Clean Printer
Nonscheduled Maintenance Procedures
Clean Shear Valve
Clean Interior of Open Sample Aspiration Probe
Unclog Open Sample Aspiration Probe
Clean Interior of Closed Sample Aspiration Needle
Unclog Closed Sample Aspiration Needle
Clean Syringes
Clean HGB Flow Cell
Clean Bar Code Reader Window
Clean Reagent Lines
Replace Open Sample Aspiration Probe
Replace Closed Sample Aspiration Needle
Replace Tubing in Sample Transfer Pump