9140181H—October 2001
Section 6
Calibration Procedures
4. Run the same Calibrator or Fresh Whole Blood sample 5 times. After each
run, observe the values displayed on the
RUN
screen for those parameters that
were calibrated and that now need confirmation. When the runs are complete,
print the appropriate QC file and transfer the mean value onto Worksheet #7.
5. Follow the instructions on this worksheet for comparing the mean value
against the reference value.
6. If the difference between the mean and reference (assay) value is within the
stated tolerance limits, then calibration has been confirmed for that
parameter. Go to step 8 below.
NOTE:
Calibration of the open mode may also be confirmed by running
commercial controls.
7. If a parameter mean is outside the tolerance limits, perform steps 4 and 5
again. If the mean is still outside the tolerance limits, call the Abbott
Hematology Customer Support Center for assistance (at 1-800-CELL-DYN
in the U.S.).
8. When Open Mode calibration has been confirmed, do one of the following:
a. If Closed Mode has not yet been calibrated, go to subsection
Auto-Cal
Procedure — Closed Mode
.
b. If Closed Mode calibration has been completed, press
[MAIN]
to return to
the
MAIN MENU
.
Auto-Cal Procedure — Closed Mode
NOTE:
Auto-Calibration is not available in the Closed Mode on the SL model. To
calibrate Closed Mode, see
Calibration Procedure — Mode to Mode
in
subsection
Enter Factor Method
.
Displaying the Auto-Cal Screen
1. If necessary, go to the
RUN
screen and press
[CHANGE SAMPLER]
to select the
Closed Mode.
2. In the
MAIN MENU
, press
[CALIBRATION]
to display the main
CALIBRATION
menu.
3. If necessary, press
[CLOSED SAMPLER]
to display the Closed Sampler
Calibration Factors.
4. Press
[PRINT]
to obtain a printout of the current Closed Sampler Calibration
Factors.
5. Press
[AUTO-CALIBRATE]
to display the
AUTO-CALIBRATION
screen.