6-24
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Calibration Procedures
Auto-Cal Method
Section 6
For example:
a. Five samples have been run and the
[ACCEPT]
is displayed after the 5th
run
b. The Factor % Diff for WOC is green, indicating it needs to be calibrated
c. The WOC result for Run #4 is highlighted, indicating it failed a reference
or precision check
d. The Mean Factor and Factor % Diff for WOC are based on only 4 runs
because Run #4 failed its internal check
e. If the operator presses the
[ACCEPT]
and
[CONFIRM ACCEPT]
keys, a new
calibrator factor will be entered for WOC but the new factor will be based
on 4, not 5, runs.
Delete a Run
If a result for a parameter is outside a predetermined limit, that result is highlighted
and not included in the Mean Factor and Factor % Diff calculations. The operator
has the option of keeping or deleting the entire run. To delete a run, do the
following:
1. Press
[DELETE A RUN]
.
2. A prompt will appear at the bottom of the screen. Type in the number of the
run to be deleted and press the Enter key.
NOTE:
Only one run can be deleted in each calibration session. The system
does not allow a deleted run to be replaced with a new run. For
example, if there are 10 runs and the operator deletes Run #5, then
calibration will be based on 9 runs (remember, a highlighted result
on a run will not be included in the Mean Factor or Factor % Diff
for that parameter).
Quit
When
[QUIT]
is pressed, the
[CONFIRM QUIT]
and
[CANCEL QUIT]
keys are displayed.
Pressing
[CONFIRM QUIT]
deletes all runs and reference values and returns the
operator to the
AUTO CALIBRATION
screen shown in Figure 6.6. The screen is ready
to receive new reference values. Pressing
[CANCEL QUIT]
returns the operator to the
current screen.
Accept
This label appears after the results of the fifth run are displayed on the screen,
allowing the operator to accept all the calibration factors shown in the lower
portion of the screen (refer to Figure 6.9). When
[ACCEPT]
is pressed, two key
labels
[CONFIRM ACCEPT]
and
[CANCEL ACCEPT]
are displayed (refer to
[CONFIRM ACCEPT]
saves the new calibration factors and
displays the
CALIBRATION LOG
screen. Pressing
[CANCEL ACCEPT]
returns the
operator to the Calibration results screen.