CELL-DYN
®
3200 System Operator’s Manual
5-87
9140181K—July 2002
Section 5
Operating Instructions
Sample Analysis on the CELL-DYN 3200SL
Introduction
This section explains how samples are run on the CELL-DYN 3200SL model in
both Open and Closed Modes. Samples should not be run until daily QC checks
have been performed and the instrument is in the READY state. For convenience,
directions for Daily Start Up and QC are given in each section. Directions for
running samples in the Open Mode are also included in each section. Samples may
be analyzed whenever the
READY
message is displayed in the Status Box on the
RUN
screen.
For Open Mode only, when specimens have not been run for one hour or more, a
background should be run immediately prior to running a patient specimen.
Samples should be well mixed (a mechanical mixer is preferred
3
) before they are
run in the Open Mode on the instrument. For specimens collected in micro-
collection devices, refer to the manufacturer’s instructions for proper mixing and
handling instructions.
The Sample Loader automatically mixes the samples immediately before aspiration.
Instrument Preparation
1. Be sure that the green READY light on the Analyzer Status Indicator Panel
is illuminated and the
READY
message is displayed in the Status Box on the
581
screen.
2. If the Status Box on the Data Station
RUN
screen displays
STANDBY
or
INITIALIZED
press
[RUN]
to initiate the automatic Start Up cycle.
3. When the automatic Start Up cycle is completed, a background count is
automatically performed and the Open Mode is selected.
4. Verify that the background counts are acceptable. If the background counts
are unacceptable, repeat the background cycle. If the counts are still
unacceptable, troubleshoot accordingly.
NOTE:
Backgrounds may be repeated by pressing the Touch Plate.
5. If the Sample Loader is to be used, be sure that at least one rack (maximum
of five) is in the load area to the right of the Tower Cover and that the Tower
Cover is in place.
6. To switch from Open Mode to Closed Mode, press the
[CHANGE SAMPLER]
key. Wait until the message
READY
is displayed in the Status Box.
7. Perform the daily Quality Control checks as directed in